Storage and Operating Conditions
Table 8. Storage and operating conditions
Property
Storage
Operating
Minimum Temperature (°C (°F)) -40 (-40) 10 (50)
Maximum Temperature (°C (°F)) 70 (158) 40 (104)
Minimum Humidity (% non-condensing) 10 30
Maximum Humidity (% non-condensing) 90 75
Minimum Atmospheric Pressure (hPa) 500 700
Maximum Atmospheric Pressure (hPa) 1060 1060
CAUTION: Allow the unit to warm to ambient temperature after transporting it from a
cold environment. If the unit was stored below 10°C (50°F), it may not function
properly.
Disposal
Return the Merlin™ PCS and the Merlin™ Antenna to St. Jude Medical at the end of their
operating lives.
The crossed-out trash can symbol marked on the programmer and the Merlin Antenna indicates
that St. Jude Medical complies with the European Union’s Waste Electrical and Electronic
Equipment (WEEE) directive. That directive calls for separate collection and disposal of hazardous
waste from electrical and electronic equipment. Sorting such waste and removing it from other
forms of waste lessens the contribution of potentially toxic substances into municipal disposal
systems and into the larger eco-system.
Accessories
Table 9. Clinical Accessories
Accessory
Model
Number
Ordered
Separately
Intended Use
5-lead ECG Cable 3626 Yes
Provide surface ECG signals for real-
time viewing on the Merlin™ PCS
3-lead ECG Cable EX3001 Yes
Provide surface ECG signals for real-
time viewing on the Merlin PCS
ECG input Cable 3617 Yes
Provide surface ECG signals for real-
time viewing on the Merlin PCS
Magnet 3630M Yes
ICDs & CRT-D: Inhibits therapy when
placed over the device
Pacemakers & CRT-Ps: Used in
battery voltage evaluation.
PSA Wand 3150 Yes
Assess the pacing and sensing
performance of the lead system prior
to pulse generator implantation, or
during invasive lead system
troubleshooting