Owner's manual

Page 7
Sonic-Pro
5.1
Power On
5.2
Run Screen
5.0 Operation
1. If power is present on Terminal T1 pins 1 & 2, the meter is always powered on.
2. If power is removed from Terminal T1 pins 1 & 2, and 4 D-Cell batteries are installed on
Terminal T1 pins 4 & 5, the meter will stay on (battery back-up). Under these conditions, the
meter can be turned on/off by pressing and holding the front panel ENTER button for three
seconds.
3. If power is only present on the battery Terminal T1 pins 4 & 5, then the meter can be
turned on/off by pressing and holding the front panel ENTER button for three seconds.
Operating time is limited when powered by batteries only.
When the S4 is factory ordered with a pipe fitting/transducer, the pipe fitting data and the
default display feature parameters are pre-configured. When power is applied and there is
water in the pipe fitting, the following RUN display screen will be displayed.
Error Code
There are three possible FAULT indicator screens.
Ÿ FAULT 11 - Transducers are not connected or there is no fluid in the pipe.
Ÿ FAULT 12 - The allowable SOS (Speed Of Sound) error set-point has been exceeded.
Ÿ FAULT 21 - The configuration data is not valid.
Flow Direction
Total Units
U.S Gallons/Min
105
1482
25
05469
U.S Gallons
.
.
0.28”
Transducer Separation
Distance
Flow Rate Units
Flow Total
Flow rate
Measured Sound Speed
Battery Life
Measurement Reliability
U.S Gallons/Min
F-11
0”
0
U.S Gallons
0.28”
XDCR Disc.
(press and hold ESC to clear)
Page 8
Sonic-Pro
5.3
Menu
Navigation
The Configuration Menu
From the RUN mode, press and release the ENTER button to enter the CONFIGURATION
menu. The following menu will be displayed:
Paddle Wheel
XDCR Settings
Meter Settings
Output Settings
Clock
Calibration
Global Config.
About
5.3.1
Selecting and
Entering Menus
Ÿ Press up arrow p or down arrow q to select menu items.
The selected item will be highlighted.
Ÿ Press right arrow u or ENTER to enter the selected menu.
Ÿ Press left arrow t to go back one menu.
Ÿ Notice the gray area. These menu items are hidden from
view. Press the up arrow p or down arrow q to display
hidden menu items.
5.3.3
Inputting Data
Ÿ Press right arrow u or left arrow t to select a digit to
modify. Pressing the left arrow t will create additional
numbers to the left.
Ÿ Press up arrow p to increase or down arrow q to
decrease the value.
Ÿ Press ENTER to confirm the modifications and return to
the menu heading.
Ÿ Press ESC to cancel the modifications and return to the
menu heading.
5.3.2
Selecting Optional
Items
Ÿ Press up arrow p or down arrow q to select optional
configuration items. The selected item will be highlighted.
Ÿ Press left arrow t or ENTER to choose the selected
optional item and return to the menu heading.
U.S. Barrels Oil
Cubic Feet
Acre Feet
Imperial Gallons
Cubic Meters
Liters
Milliliters
FR Custom V/Gal
U.S. Gallons
Ounces
U.S. Barrels Liq.
Enter Value
9.1000
0
5.3.4
Save Configuration
Ÿ Press left arrow t to exit to the main menu.
Ÿ Press down arrow q to Highlight Save Config.
Ÿ Press right arrow u or ENTER to save the configuration
changes.
Ÿ Press ENTER to confirm the configuration changes have
been saved.
Ÿ Press up arrow p to RUN.
Ÿ Press right arrow u or ENTER to enter the RUN mode,
Configuration
XDCR Settings
Save Config.
Configuration
Configuration Saved!
OK
XDCR Settings
Save Config.
Configuration
5.1.1
Battery Powered
operation
The following meter features are disabled during battery operation:
Ÿ Communications Module
Ÿ Analog output
Ÿ Pulse Frequency Output
Continuous Running
On / Sleep Time (seconds)
10 / 10
128 Hr72 Hr 310 Hr 377 Hr 450 Hr
10 / 100 5 / 100 3 / 268
Ÿ Press the ENTER button to enter the menu system.
Ÿ Press the ESC button to clear the flow total to zero (if enabled in configuration).
Ÿ Press the up arrow p to toggle the flow rate to linear Feet/Sec., Meters/sec and back to
the configured user units of measure.
5.2.1
Fault Screen
While the batteries pack is provided primarily as back-up power, the S4 meter can be
operated solely with 4 D-cell batteries (not included). The meter can be configured to sleep
between measurements in order to conserve power. The approximate battery life is .....