User's Manual

14 1514 15
GUIDANCE AND MANUFACTURER’S DECLARATION - EMISSIONS
The Armband is intended for use in the electromagnetic environment specified below. The customer or user
of the Armband should ensure that it is used in such an environment.
IMMUNITY TEST
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
ELECTROMAGNETIC ENVIRONMENTGUIDANCE
Portable and mobile communications equipment
should be separated from Armband by no less than
the distances calculated/listed below:
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P) 80 to
800 MHz
D=(7/EI)(Sqrt P)
800 MHz to 2.5 GHz
Where P is the max power in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as determined
by an electromagnetic site survey, should be less than
the compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment
containing a transmitter.
RECOMMENDED SEPARATION DISTANCES FOR THE PRODUCT
The Armband is intended for use in the electromagnetic environment specified below. The customer or user
of the Armband can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF Communications Equipment and the Armband as recommended below, according to
the maximum output power of the communications equipment.
MAX OUTPUT POWER
(WATT S)
SEPARATION (M) 150KHZ
TO 80MHZ
D=(3.5/V1)(SQRT P)
SEPARATION (M) 80 TO
800MHZ
D=(3.5/V1)(SQRT P)
SEPARATION (M) 800MHZ
TO 2.5GHZ
D=(7/E1)(SQRT P)
0.01
0.1
1
10
100
0.1166
0.3689
1.1666
3.6893
11.6666
0.1166
0.3689
1.1666
3.6893
11.6666
0.2333
0.7378
2.3333
7.37 8 6
23.3333