User's Manual

12 1312 13
USER ENVIRONMENT
1.5m 1.5m
2.5m
1.5m
Computer and Wireless
Communicator
Diagram not to scale.
GUIDANCE AND MANUFACTURER’S DECLARATION - EMISSIONS
The Armband is intended for use in the electromagnetic environment specified below. The customer or user
of the Armband should ensure that it is used in such an environment.
EMISSIONS TEST
RF Emissions CISPR 11
Harmonics IEC 6100-3-2
Flicker IEC 6100-3-3
COMPLIANCE
Class B, Group 1
N/A
N/A
ELECTROMAGNETIC ENVIRONMENTGUIDANCE
The Armband uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment
The Armband is suitable for use in all establishments,
including domestic, and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes
GUIDANCE AND MANUFACTURER’S DECLARATION - IMMUNITY
The Armband is intended for use in the electromagnetic environment specified below. The customer or user
of the Armband should ensure that it is used in such an environment.
IMMUNITY TEST
ESD IEC 61000-4-2
EFT IEC 61000-4-4
Voltage Dips/
Dropout IEC
61000-4-11
Power Frequency
50/60H
Magnetic Field IEC
61000-4-8
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
±6kV Contact
±8kV Air
±2kV Mains
±1kV I/Os
±1kV Differential
±2kV Common
>95% Dip for
0.5 Cycles
60% Dip for 5
Cycles
30% Dip for 25
Cycles
>95% Dip for 5
Seconds
3A/m
±6kV Contact
±8kV Air
N/A
N/A
N/A
3A/m
ELECTROMAGNETIC ENVIRONMENTGUIDANCE
Floors should be wood, concrete, or ceramic tile. If
floors are synthetic, they r/h should be at least 30%
Mains power quality should be that of a typical
commercial or hospital environment
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
Armband requires continued operation during power
mains interruptions, it is recommended that the
Armband be powered from an uninterruptible power
supply or battery.
Power frequency magnetic fields should be that of a
typical commercial or hospital environment.