Boekel Round Tissue Bath Models 14792 and 14792-2 Operating Instructions N2400142 Rev.
CONTENTS 1. 2. 3. 3.1 3.2 4. 4.1 4.2 4.2 5. 7. 8. 8.1 8.2 9. 10. Page Safety .............................................................................................................................. 2 Product Information ........................................................................................................ 3 Assembly......................................................................................................................... 3 Unpacking ....................................
1. Safety The following symbols marked on the equipment mean: Caution: Read these operating instructions fully before use and pay particular attention to sections containing this symbol. Attention: Suivre attentivement les instructions avant l’usage et prêtez une attention particulière aux sections comportant ce symbole. Caution: Surfaces can become hot during use. Attention: Les surfaces peuvent devenir brûlantes pendent l’usage.
2. Product Information The Boekel Round Tissue Bath is designed as an economical and efficient tissue floatation bath for use in mounting tissue specimens to glass slides. The Round Tissue Bath has an 8.5” (21.6cm) diameter by 2” (5cm) deep-water chamber with a 1.25” (3.2cm) wide rim for slide warming and drying. The Tissue Bath has a water temperature range of ambient + 10°C to 65°C under normal operating conditions. 3. Assembly 3.
4.2 Filling the Bath Fill the bath with water. Always keep at least 3/4” (1.9cm) of water in the bath. Do not let the bath run dry as heater damage may result. 5. Fault Diagnosis Symptom 1. Unit does not operate 2. Chamber temperature does not rise when expected 3. Temperature continues to rise when not expected 7. Possible Cause a. Unit not switched on b. Unit not plugged into power supply c. Fuses blown d. Power supply failure Action Required a. Switch on b. Plug in, switch on a.
8. Maintenance and Service All Boekel laboratory products are designed to comply with IEC1010-1. maintenance is required. No routine 8.1 Cleaning Disengage power cord prior to cleaning. The outer casing may be cleaned with water and a damp cloth. Do not submerge or immerse the Round Tissue Bath in water. Before using any cleaning or decontamination method except those recommended by the manufacturer, users should check that the proposed method would not damage the equipment. 8.
Boekel Scientific 855 Pennsylvania Blvd. Feasterville, PA 19053 PHONE: (215) 396-8200 or (800) 336-6929 FAX: (215) 396-8264 e-mail: boekel-info@boekelsci.com website: www.boekelsci.