Boekel Bambino™ Hybridization Oven Models 230301 and 230301-2 Operating Instructions N2400204 Rev 0 11/03
CONTENTS 1. 2. 3. 3.1 3.2 4. 4.1 4.2 4.3 4.4 4.5 4.6 5. 5.1 5.2 6. 7. 8. 8.1 8.2 9. 10. Page Safety ................................................................................................................ 2 Product Information........................................................................................... 3 Assembly.......................................................................................................... 3 Unpacking ........................................................
1. Safety The following symbols marked on the equipment mean: Caution: Read these operating instructions fully before use and pay particular attention to sections containing this symbol. Caution: Surfaces can become hot during use. Always observe the following safety precautions: Use only as specified by the operating instructions or the intrinsic protection may be impaired. After transport or storage in humid conditions, dry out the unit before connecting it to the supply voltage.
2. Product Information The Boekel Hybridization Oven Model 230301 & 230301-2 hybridization oven will hold up to four 50ml plastic centrifuge tubes. The Oven can also be used with other size tubes using the optional carousel. The Boekel Hybridization Oven utilizes an adjustable speed Carousel and forced air circulation coupled with proportional (PID) temperature control that combine to provide unsurpassed temperature performance and the most consistent conditions from procedure to procedure.
button ‘▲’, the set temperature value is raised. When all buttons are released, the actual chamber temperature is displayed. In the event of power loss, the Temperature Controller retains the last set temperature value. 4.3 Loading Carousel Warning! Sharp rotating parts. Do not attempt to add or remove tubes until the motor has stopped. The motor can be stopped by turning the power switch to the “Off” position or by depressing the “Pause” button located on the back of the unit, (see figure 2).
5.2 Clips for Multi-Tube Carousel (22 clips per bag) four clips for 50ml tubes six clips for 15ml tubes twelve clips for 1.5/2.0ml tubes Catalog Number 230350 6. Fault Diagnosis Symptom 1. Unit does not operate 2. Chamber temperature does not rise when expected 3. Temperature continues to rise when not expected 4. No rotation, LEDs illuminated 7. Possible Cause a. Unit not switched on b. Unit not plugged into power supply c. Fuses blown d. Power supply failure a.
Heating Rate Speed of Rotation Ambient to 70°C within 25 minutes 6 to 20 RPM Note: A minimum of at least two tubes are needed for optimum temperature stability. 8. Maintenance and Service The Model 230301 & 230301-2 Oven is designed to comply with IEC1010-1. No routine maintenance is required. 8.1 Cleaning Disengage power cord prior to cleaning. If a spill occurs, use appropriate clean up procedures as required for radiation or biohazard control. The outer casing can be cleaned with a damp cloth.
Temperature Display Heater Lamp Figure 1 Pause Button Speed Control Knob Power Switch I Figure 2 7 Fuse Drawer