Boekel Slide Moat Model 240000 & 240000-2 Operating Instructions N2400080 Rev.
Table of Contents 1.0 2.0 3.0 Safety ............................................................................................3 Product Information ...................................................................4 Assembly.......................................................................................4 4.0 Operation .....................................................................................4 5.0 6.0 7.0 8.0 3.1 Unpacking the Unit ..................................................
Figure 1 Front Panel Heat Temperature Display Indicator MODEL 240000 POWER TEMPERATURE Power Switch Figure 2 Rear of Unit Fuse Drawer 2
1.0 Safety The following symbols marked on the equipment mean: Caution: Read these operating instructions fully before use and pay particular attention to sections containing this symbol. Attention: Suivre attentivement les instructions avant l’usage et prêtez une attention particulière aux sections comportant ce symbole. Caution: Surfaces can become hot during use. Attention: Les surfaces peuvent devenir brûlantes pendant l’usage.
2.0 Product Information The Boekel Slide Moat is designed to provide a heating platform and chamber capable of maintaining either dry or highly humid conditions for In-Situ Hybridization applications. Uses include in-situ RNA amplification, reverse transcription reactions, hybridization and DNA melt. The Boekel Slide Moat can hold up to 30 standard microscope slides.
The Power Switch controls power to the unit. The Heat Indicator is on continuously while the oven is heating up. As the required temperature is approached, it starts to flash. When the unit is controlling at the set temperature, the heater lamp flashes intermittently. The Temperature Display shows the chamber temperature in degrees Celsius. 4.2 Preparing the unit Switch the unit on (I) and set the PID Controller to the desired operating temperature.
Temperature Range Temperature Setting Range Stability Uniformity Temperature Display Resolution Supply Ratings Power Rating Heating Rate 6.0 Ambient + 5°C to 100°C 10.0°C to 100.0°C ±0.2°C throughout operating temperatures ±0.35°C at 37°C; ±0.7°C at 65°C 0.1°C Model 240000: 115 VAC, 1.74 A, 60 Hz Model 240000-2: 230 VAC, 0.87 A, 50/60 Hz Model 240000: 200 W Model 240000-2: 200 W Ambient to 65°C in 20 minutes Fault Diagnosis Symptom Unit does not operate Possible Cause Action Required a.
.0 Maintenance and Service All Boekel laboratory products are designed to comply with IEC1010-1. No routine maintenance is required. 8.1 Cleaning Disengage power cord prior to cleaning. If a spill occurs, use appropriate clean-up procedures as required for radiation or biohazard control. The outer casing can be cleaned with a cloth dampened with water. Do not immerse the unit in water. 8.2 Replacement of Fuses There are two supply fuses located in the fuse drawer.
Boekel Scientific 855 Pennsylvania Boulevard Feasterville, PA 19053 Phone: (215) 396-8200 or (800) 336-6929 Fax: (215) 396-8264 e-mail: boekel-info@boekelsci.