Boekel MiniFridge II Models 260009 and 260009-2 Operating Instructions N2400149 Rev.
CONTENTS Page 1. Safety .............................................................................................................................. 2 2. Product Information ........................................................................................................ 3 3. Assembly......................................................................................................................... 3 3.1 Unpacking ................................................................................
1. Safety The following symbols marked on the equipment mean: Caution: Read these operating instructions fully before use and pay particular attention to sections containing this symbol. Attention: Suivre attentivement les instructions avant l’usage et prêtez une attention particulière aux sections comportant ce symbole. Always observe the following safety precautions: • • • • • • • • • • • • • Use only as specified by the operating instructions or the intrinsic protection may be impaired.
2. Product Information The Boekel MiniFridge II cold block incubator is a convenient and reliable way for labs to incubate or store samples at temperatures from ambient down to 18°C below ambient, typically 4°C . The “dry cold” of the MiniFridge II makes it the ideal substitute for those messy ice buckets. The unit’s simple “press to set” controller makes it perfect for maintaining the 14°C used in ligation reactions or the 17°C used to store oocytes.
4. 4.1 Operation Controls and Displays 2. Cooling Indicator LED 1. Power key 3. Temperature Display 4. Set function key 5. Up and Down arrow keys 1. Power Key ( ) – Controls Power to the unit 2. Cooling Indicator Light-Green LED indicated cooling Function 3. Temperature Display- Displays the current well temperature of the unit. 4. Set (S) function Key- is used to adjust the temperature of the unit 5. Up () and Down () arrow keys-are used to adjust the set temperature 4.
the Cooling Function. A blinking indicator LED shows that the system is nearing the temperature set point and will stabilize to the set temperature. 4.4 Operating the MiniFridge II The MiniFridge II is rated to work at temperatures between ambient and to 18°C below ambient, typically 4°C.
5.10 Block – 35 x 0.4 ml tubes Boekel Part Number 110047 5.11 Block – 48 x 0.2 ml tubes Boekel Part Number 110048 5.12 Block – 24 x 2.0ml Dolphin tubes Boekel Part Number 110099 6. Fault Diagnosis Symptom 1. Unit does not operate 2. Well temperature does not rise when expected 3. Well temperature continues to rise when not expected 4. Well temperature does not cool when not expected 5. Well temperature continues to cool when not expected 7. Possible Cause a. Unit not powered on b.
This equipment is intended for indoor use and will meet its performance figures within the ambient temperature range of 10°C to 35°C, with maximum relative humidity of 80% (noncondensing). Installation Category II (transient voltages). Pollution Degree 2 in accordance with IEC 664. Suitable for operation at altitudes of up to 6500 feet. Specifications: Temperature Range: Supply Voltage Range: Power Rating: 8.
. Warranty Fuse Drawer Figure 2 When used in laboratory conditions and according to these operating instructions, Boekel warrants this product to be free of defective material and workmanship for a period of two years from the date of manufacture. The liability of Boekel for any defective equipment during the warranty period shall be limited to the repair of such equipment or replacement thereof without charge for parts or labor. 10.