This equipment is intended for indoor use and will meet its performance figures within the
ambient temperature range of 10°C to 35°C, with maximum relative humidity of 80% (non-
condensing). Installation Category II (transient voltages). Pollution Degree 2 in accordance
with IEC 664. Suitable for operation at altitudes of up to 6500 feet.
Specifications:
Temperature Range: ambient to 18°C below ambient, typically 4°C
Supply Voltage Range: 115V +/- 10%, 50/60 Hz
230V +/- 10%, 50/60 Hz
Power Rating: Model 260009: 110 W, 1.3 A
Model 260009-2: 110 W, 0.65A
8. Maintenance and Service
All Boekel laboratory products are designed to comply with IEC1010-1. No routine
maintenance is required.
8.1 Cleaning
Disengage power cord prior to cleaning. The outer casing may be cleaned with water and
a damp cloth. Do not submerge or immerse the MicroCooler II in water. Abrasive
materials such as paper towels may scratch the well cover or the plastic body. Do not use
solvents to clean the well cover or the plastic body. Before using any cleaning or
decontamination method, users should check that the proposed method would not damage
the equipment.
8.2 Replacement of Fuses
There are two supply fuses located in the fuse drawer. To replace the fuses:
Disconnect the unit from the power supply.
Remove the plug from the socket in the back of the unit.
Pull back on the fuse drawer (see Figure 2).
Pull out the fuse drawer.
Check and replace with the correct fuses if necessary. The fuses must be
5mm x 20mm quick acting, rated 250V.
Model 260009: 2.0AF
Model 260009-2: 1.0AF
Push the fuse drawer back in. Reconnect unit to the power supply.