Operating instructions

28 GMS810-/GMS811-FIDOR · Operating Instructions · 8013911 V1-0 · © SICK AG
Installation and Connection
Subject to change without notice
X5 1 relay contact 5
normally open
DO5 N3M04DO01 (DO04) Measuring Measured value OK
2 relay contact 5
common
3 relay contact 5
normally closed
4 relay contact 6
normally open
DO6 N3M04DO02 (DO04) Adjustment Adjustment running
5 relay contact 6
common
6 relay contact 6
normally closed
7 relay contact 7
normally open
DO7 N3M04DO03 (DO04) Maintenance mode Maintenance mode active
8 relay contact 7
common
9 relay contact 7
normally closed
10 relay contact 8
normally open
DO8 N3M04DO04 (DO04) Measuring range ID AO3
Identification of active measuring
range from analog output AO3
11 relay contact 8
common
12 relay contact 8
normally closed
X7 1
ground GND
2
3 (+) analog input 1
(0 … 20 mA)
AI1
N3M07AI01 (AI02)
Reserved Not used
4 (+) analog input 2
(0 … 20 mA)
AI2
N3M07AI02 (AI02)
Reserved Not used
5 (–) analog output 1
AO1 N3M05AO01 (AO02) Measured value
Output of measured value in the
set unit and set measuring range.
Parameter setting via FIDOR
6 (+) analog output 1
(0/2/4 … 20 mA)
7 (–) analog output 2
AO2 N3M05AO02 (AO02) Measured value
Output of measured value in the
set unit and set measuring range.
Parameter setting via FIDOR
8 (+) analog output 2
(0/2/4 … 20 mA)
9 (–) Analog output 3
AO3 N3M06AO01 (AO02) Set measured value
The measured value output is set
to two measuring ranges 15/30
(automatic switchover). Parame-
ter setting via BCU/SCU
10 (+) analog output 3
(0/2/4 … 20 mA)
11 (–) Analog output 4
AO4 N3M06AO02 (AO02) Set measured value
The measured value output is set
to two measuring ranges 15/30
(automatic switchover). Parame-
ter setting via BCU/SCU
12 (+) analog output 4
(0/2/4 … 20 mA)
Terminal Pin Function Name Significance Explanation