5.2.5.2 Logbook
Menu: FIDOR/Diagnostic/Logbook
5.2.5.3 Counters
Menu: FIDOR/Diagnostic/Counters
This menu shows the operating hours counter.
5.2.5.4 Hardware
Menu: FIDOR/Diagnostic/Hardware
This menu shows current FIDOR-internal values and states.
Menu item Significance
Operating Operating hours [hours] (“operation” means: Device is on).
Menu item Submenu Significance
pA-amplifier pA-amplifier (measurement amplifier)
– Meas. val. – Actual measured value
– Range – Actual measuring range
– Unit – Actual unit
– Meas.out – Actual output current for actual measured value
– mgC lin. – Actual measured value in mgC linearized
– mgC raw – Actual measured value in mgC (raw value)
– pA – Actual current of pA-amplifier in picoampere
– State – State of pA measurement amplifier (OK or FAIL)
– Offset – Offset of pA-amplifier in counts
– Offset – Offset of pA-amplifier (internal unit)
Digital input – Compress air – Compressed air pressure switch (ON/OFF)
ON: Compressed air available
– Fuel gas – Fuel gas pressure switch (ON/OFF)
ON: Fuel gas available
– Test gas – Test gas pressure switch (ON/OFF)
ON: Test gas available
– Int. ready – Internally ready/not ready
– SD-card – SD-card is inserted/not inserted
– CAN-Adr. – Internal CAN address of the FIDOR
Digital output – Ignition plug – ON (ignited), OFF (not ignited)
– SOV zero gas – Zero gas solenoid valve ON (open)/OFF (closed)
–SOV reference
gas
– Reference gas solenoid valve ON (open)/OFF (closed)
.4.2.2.1
FIDOR:MEASURING
Back
Logbook
.\FIDOR/Diagnostic/Logbook
The currently existing messages are shown in this menu
(Logbook SOPAS ET).
1 / 3 09-01-01 08:25
Failure
On #1: Pressure
● Date of occurrence (yy-mm-dd).
Time of occurrence (hh:mm).
● Current message / number of existing messages
● Error message (error number and plain text).
(List of messages
p. 66, §8.6).