Operating instructions

Operation
GMS810-/GMS811-FIDOR · Operating Instructions · 8013911 V1-0 · © SICK AG 43
Subject to change without notice
5.2.5.2 Logbook
Menu: FIDOR/Diagnostic/Logbook
5.2.5.3 Counters
Menu: FIDOR/Diagnostic/Counters
This menu shows the operating hours counter.
5.2.5.4 Hardware
Menu: FIDOR/Diagnostic/Hardware
This menu shows current FIDOR-internal values and states.
Menu item Significance
Operating Operating hours [hours] (“operation” means: Device is on).
Menu item Submenu Significance
pA-amplifier pA-amplifier (measurement amplifier)
Meas. val. Actual measured value
Range Actual measuring range
Unit Actual unit
Meas.out Actual output current for actual measured value
mgC lin. Actual measured value in mgC linearized
mgC raw Actual measured value in mgC (raw value)
pA Actual current of pA-amplifier in picoampere
State State of pA measurement amplifier (OK or FAIL)
Offset Offset of pA-amplifier in counts
Offset Offset of pA-amplifier (internal unit)
Digital input Compress air Compressed air pressure switch (ON/OFF)
ON: Compressed air available
Fuel gas Fuel gas pressure switch (ON/OFF)
ON: Fuel gas available
Test gas Test gas pressure switch (ON/OFF)
ON: Test gas available
Int. ready Internally ready/not ready
SD-card SD-card is inserted/not inserted
CAN-Adr. Internal CAN address of the FIDOR
Digital output Ignition plug ON (ignited), OFF (not ignited)
SOV zero gas Zero gas solenoid valve ON (open)/OFF (closed)
–SOV reference
gas
Reference gas solenoid valve ON (open)/OFF (closed)
.4.2.2.1
FIDOR:MEASURING
Back
Logbook
.\FIDOR/Diagnostic/Logbook
The currently existing messages are shown in this menu
(Logbook SOPAS ET).
1 / 3 09-01-01 08:25
Failure
On #1: Pressure
Date of occurrence (yy-mm-dd).
Time of occurrence (hh:mm).
Current message / number of existing messages
Error message (error number and plain text).
(List of messages
p. 66, §8.6).