5.2.7 Adjustment
Menu: FIDOR/Adjustment
5.2.7.1 Adjustment results
Menu: FIDOR/Adjustment/Adjustment results
This menu shows the adjustment results.
5.2.7.2 Zero + reference
Menu: FIDOR/Adjustment/Zero + reference
Start an automatic zero and reference point adjustment.
Menu item Significance
Act. Date of actual zero point adjustment.
Zero Time of actual zero point adjustment.
ZDrift abs. Absolute zero point drift since last adjustment. Limit value as default. Message cre-
ated when exceeded.
ZDrift rel. Relative zero point drift since last adjustment. Limit value as default. Message cre-
ated when exceeded.
Act. Date of actual reference point adjustment.
Ref. Time of actual reference point adjustment.
RDrift abs. Absolute reference point drift since last adjustment. Limit value as default. Mes-
sage created when exceeded.
RDrift rel. Relative reference point drift since last adjustment. Limit value as default. Message
created when exceeded.
Last Date of previous zero point adjustment.
Zero Time of previous zero point adjustment.
ZDrift abs. Absolute drift since factory adjustment.
ZDrift rel. Relative drift since previous zero point adjustment.
Last Date of previous reference point adjustment.
Ref. Time of previous reference point adjustment.
RDrift abs. Absolute drift since factory adjustment.
RDrift rel. Relative drift since previous reference point adjustment.
Menu item Significance
Start adj.
1 Select menu item.
2 Start zero and reference point adjustment with <Set>.
Process Actual device process (e.g. “ADJUSTMENT”).
M. value Actual measured value
3
p. 48, §5.2.7.1
p. 48, §5.2.7.2
p. 49, §5.2.7.3
p. 49, §5.2.7.4
p. 49, §5.2.7.5
p. 49, §5.2.7.6
p. 49, §5.2.7.7
p. 50, §5.2.7.8
FIDOR:MEASURING
Back
Enter
Adjustment
.\FIDOR
1 Adjustment results
2 Zero + reference
3 Zero
4 Reference
5 Abort adjustment
6 Adjustment time
7 Restart cycle
8 Drift reset