Operating instructions

52 GMS810-/GMS811-FIDOR · Operating Instructions · 8013911 V1-0 · © SICK AG
Operation
Subject to change without notice
5.2.8.3 Test gas
Menu: FIDOR/Maintenance/Test gas
Zero and reference gas can be activated in this menu.
An adjustment is not performed.
5.2.8.4 Standby
Menu: FIDOR/Maintenance/Standby
The operating state "Standby" is activated/deactivated in this menu.
Standby: The flame is switched off, the device is heated further and purged with zero gas.
5.2.8.5 Configurations
Menu: FIDOR/Maintenance/Configurations
This menu is used to save and load the configuration.
Menu item Subitem Significance
Zero gas Zero gas set To switch on the zero gas: <Set>.
Duration Selection: “Infinite” or “Limited”.
Time limit Enter with <Set>, <>, <>, <Save>.
Unit: [s].
Deactivation is then automatic.
Process Actual device process (e.g. “MEASURING”).
M.value Measured value in selected unit. Is continuously updated.
M.value Is continuously updated.
Reference gas Ref. gas To switch on the reference gas: <Set>.
Duration Selection: “Infinite” or “Limited”.
Time limit Enter with <Set>, <>, <>, <Save>.
Unit: [s].
Deactivation is then automatic.
Process Actual device process (e.g. “MEASURING”).
M.value Measured value in selected unit. Is continuously updated.
M.value Is continuously updated.
Test gas off Test gas off To switch off the reference gas: <Set>.
Process Actual device process (e.g. “MEASURING”).
M.value Measured value in selected unit. Is continuously updated.
M.value Is continuously updated.
Menu item Significance
Set standby
Select the desired state and confirm with <Set>.
Exit standby
Menu item Significance
Save user setup Save the user setup.
Select the desired action and con-
firm with <Set>.
Load user setup Load the saved user setup.
Load factory setup Load the factory setup.