INSTRUCTION MANUAL 4540 SERIES RF POWER METER This manual covers instrument serial #s: 11001 and higher Wireless Telecom Group 25 EASTMANS ROAD, PARSIPPANY, NJ 07054 Telephone: 973-386-9696 Fax: 973-386-9191 Email: boonton@boonton.com Web: www.wtcom.
Boonton 4540 Series RF Power Meter INSTRUCTION MANUAL, 4540 SERIES RF POWER METER Revision date 10/04/2012 © Copyright in 2005-2012, by BOONTON Electronics, a subsidiary of the Wireless Telecom Group, Inc. Parsippany, NJ, USA. All rights reserved. P/N 98406100A This manual covers instrument serial numbers: 11001 and higher. The 4540 Application software used in this product is licensed by Boonton Electronics, a subsidiary of the Wireless Telecom Group, Inc.
Boonton 4540 Series RF Power Meter SAFETY SUMMARY The following general safety precautions must be observed during all phases of operation and maintenance of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Boonton Electronics assumes no liability for the customer’s failure to comply with these requirements.
Boonton 4540 Series RF Power Meter SAFETY SYMBOLS This safety requirement symbol (located on the rear panel) has been adopted by the International Electro-technical Commission, Document 66 (Central Office) 3, Paragraph 5.3, which directs that an instrument be so labeled if, for the correct use of the instrument, it is necessary to refer to the instruction manual. In this case it is recommended that reference be made to the instruction manual when connecting the instrument to the proper power source.
Boonton 4540 Series RF Power Meter 1. General Information .............................................................................................. 1-1 1.1 Organization.......................................................................................................... 1-1 1.2 Description ............................................................................................................ 1-2 1.3 Features .....................................................................................
Boonton 4540 Series RF Power Meter 4. Operation ................................................................................................................ 4-1 4.1 Manual Operation ................................................................................................. 4-1 4.2 Control Menus ...................................................................................................... 4-1 4.3 Numerical Data Entry .........................................................................
Boonton 4540 Series RF Power Meter 4.5.39 4.5.40 4.5.41 4.5.42 4.5.43 4.5.44 4.5.45 4.5.46 4.5.47 4.5.48 4.5.49 4.5.50 4.5.51 4.5.52 4.5.53 4.5.54 4.5.55 System > I/O Config > Multi IO Menu .................................................... 4-63 System > I/O Config > Multi IO > IO Mode Status Out Menu ................ 4-64 System > I/O Config > Multi IO > IO Mode Rec Out Menu ................... 4-65 System > I/O Config > Multi IO > Rec Out > Rec Limits Menu .............
Boonton 4540 Series RF Power Meter *SRE .................................................................................................................... 5-8 *STB? .................................................................................................................. 5-9 *TRG.................................................................................................................... 5-9 *TST? ...............................................................................................
Boonton 4540 Series RF Power Meter 5.6.5 FETCh Queries ............................................................................................ 5-23 FETCh:ARRay:AMEAsure:POWer? ................................................................ 5-23 FETCh:ARRay:AMEAsure:STATistical? ......................................................... 5-23 FETCh:ARRay:AMEAsure:TIMe? ................................................................... 5-24 FETCh:ARRay:CW:POWer? ..........................................
Boonton 4540 Series RF Power Meter MEMory:SNSR:LFLIN? ................................................................................... 5-36 MEMory:SNSR:MESSage? ............................................................................... 5-36 MEMory:SNSR:TEMPCOMP? ......................................................................... 5-36 MEMory:SYS:LOAD ........................................................................................ 5-37 MEMory:SYS:STORe ...................................
Boonton 4540 Series RF Power Meter READ:MARKer:CURsor:POWer? ................................................................... 5-50 READ:MARKer:RATio? .................................................................................. 5-50 READ:MARKer:RDELTa? ............................................................................... 5-51 READ:MARKer:RRATio? ................................................................................ 5-51 5.6.12 SENSe Subsystem......................................
Boonton 4540 Series RF Power Meter STATus:QUEStionable:EVENt? ....................................................................... 5-67 STATus:QUEStionable:NTRansition ................................................................ 5-67 STATus:QUEStionable:PTRansition ................................................................ 5-67 STATus:QUEStionable:CALibration:CONDition? .......................................... 5-68 STATus:QUEStionable:CALibration:ENABle .......................................
Boonton 4540 Series RF Power Meter TRIGger:POSition ............................................................................................. 5-80 TRIGger:SLOPe ................................................................................................ 5-80 TRIGger:SOURce .............................................................................................. 5-81 TRIGger:VERNier ............................................................................................. 5-81 5.6.
Boonton 4540 Series RF Power Meter 9.3 Optional Configurations ....................................................................................... 9-1 Other Options ........................................................................................................... 9-1 9.4 Specifications ........................................................................................................ 9-2 RF Characteristics ....................................................................................
Boonton 4540 Series RF Power Meter 1. General Information This instruction manual provides you with the information you need to install, operate and maintain the Boonton 4540 Series RF Power Meter. Section 1 is an introduction to the manual and the instrument. Throughout this manual, the designation ―4540‖ will be used to mean the 4540 Series RF Power Meter, which includes both the single-channel Model 4541 and the dual-channel Model 4542. 1.
Boonton 4540 Series RF Power Meter 1.2 Description The 4540 Series RF Power Meter is the latest generation of BOONTON RF Peak Power Meters and Analyzers, including Models 4400, 4500, 4400A, 4500A, 4500B and 4530. The 4540 Series, when operated with Boonton peak and CW power sensors, provides one of the most versatile power measuring systems available, with capability to make over 20 different power related measurements on captured signals.
Boonton 4540 Series RF Power Meter 1.3 Features Software. A 32-bit control computer running a real-time operating system provides display, I/O and system memory functions for the instrument. The instrument also contains a dedicated floating-point DSP, FPGA and memories to perform the measurement functions. Software updates are easily made using the rear-panel Ethernet port. Auto-Setup.
Boonton 4540 Series RF Power Meter Dual Independent Channels. When equipped with the optional second measurement channel, the instrument can display two pulsed RF signals simultaneously or one pulsed signal and one CW signal. Each channel is calibrated and all channel parameters are channel-independent. Balanced Diode Sensors. The balanced diode sensor configuration of most Boonton peak and CW power sensors provides high sensitivity and even-order harmonic suppression. Low VSWR minimizes mismatch errors.
Boonton 4540 Series RF Power Meter 1.4 Accessories Optional 4540 accessories that can be ordered from Boonton Electronics. Table 1-1 Accessories for the 4540 Series Selection Part Number Description 56810400A 96101301A 98601300C Line Cord (US) Fuse Kit Manual CD, Boonton Measurement Instruments (CD-ROM) 54600000A 95403001A 95403003A 95105501A 95600005A 95600010A 95600020A 95600025A 95600050A 95109101A 95109102A 95109001A 41-2A 41-2A/10 41-2A/20 41-2A/50 41-2A/100 98406100A Fuse, USA (1.
Boonton 4540 Series RF Power Meter 1.6 Specifications Performance specifications for the 4540 Series are listed in Table 1-2. Performance specifications for all Boonton power sensors are found in the Boonton Sensor Manual, which may be ordered as Boonton p/n 98501900J.
Boonton 4540 Series RF Power Meter Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subject to change without notice) TRIGGER Trigger Mode: Normal, Auto, Auto Pk-to-Pk, Free Run Trigger Source: Channel 1 (internal) Channel 2 (internal) External Trigger Internal Trigger Level Range: -40 to +20 dBm (sensor dependent) External Trigger Level Range: ±5 volts, ±50 volts with 10:1 divider probe.
Boonton 4540 Series RF Power Meter Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subject to change without notice) PULSE MODE OPERATION Acquisition Mode: Discontinuous triggered sample acquisition Trace Display: Power versus time swept trace (rolled trace for slow timebases) Trace Averaging: 1 to 16384 samples per sweep data point, exponential Markers (vertical cursors): Settable in time relative to the trigger position.
Boonton 4540 Series RF Power Meter Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subject to change without notice) CALIBRATION SOURCE Internal Calibrator Operating Modes: Off, On CW Frequency: 50.025 MHz ± 0.1% Level Range: -60 to +20 dBm Resolution: 0.1 dB RF Connector: Type N Source VSWR: 1.05 (reflection coefficient = 0.024) Accuracy, 0C to 20C, NIST traceable: At 0 dBm: ±0.055 dB (1.27%) +20 to -39 dBm: ±0.075 dB (1.74%) -40 to -60 dBm: ±0.105 dB (2.
Boonton 4540 Series RF Power Meter Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subject to change without notice) PHYSICAL AND ENVIRONMENTAL CHARACTERISTICS Case Dimensions: 8.4W x 3.5H x 13.5D inches (21.3 x 8.9 x 34.3 cm), Half-rack width, 2U height Weight: 7.7 lbs (3.5kg) Power Requirements: 90 to 260 VAC, 47 to 63 Hz, 50W (70VA) maximum. Fuse 1.
Boonton 4540 Series RF Power Meter 2. Installation This section contains unpacking and repacking instructions, power requirements, connection descriptions and preliminary checkout procedures. 2.1 Unpacking & Repacking The 4540 Series is shipped complete and is ready to use upon receipt. Figure 2-1 shows you the various pieces included in the packaging and the order in which they are loaded into the container. Actual details may vary from the illustration.
Boonton 4540 Series RF Power Meter Table 2-1 4540 Series Packing List INSTRUMENT (See also Table 1-1) 4540 Series RF Power Meter Line Cord Fuse Kit Boonton Instruction Manual CD SENSOR(S) (packaged separately) Sensor(s) Sensor Cable(s) Type N to SMA Adapter (if required) BOONTON Sensor Manual CD For bench-top use, choose a clear, uncluttered area. Ensure that there is at least 2" of clearance at the fan air intake on the rear panel and the exhaust vents on the side panels.
Boonton 4540 Series RF Power Meter Trigger Most triggered applications can use the RF signal applied to the sensors for triggering. For measurements requiring external triggering, connect the external trigger signal to the EXT TRIGGER BNC connector on the rear panel. Remote If the instrument is to be operated remotely using the GPIB (IEEE-488) bus, connect the instrument to the bus using the rear panel GPIB connector and appropriate cable.
Boonton 4540 Series RF Power Meter 6. On the front panel, press the Menu key followed by the More softkey, then on the softkeys, press System > Reports > Configuration REPORT. A display similar to Figure 2-3 should appear. Figure 2-3. System > Configuration Report Display 7. Verify that the instrument serial number and model number shown are correct, and the correct number of channels (single or dual) appears.
Boonton 4540 Series RF Power Meter 3. Getting Started This chapter will introduce the user to the 4540 Series. The chapter will identify objects on the front and rear panels, identify display organization, list the initial configuration of the instrument after reset, demonstrate how to calibrate the sensors, and provide practice exercises for front panel operation. For additional information you should see Chapter 4 "Operation." 3.1 Organization Subsection 3.
Boonton 4540 Series RF Power Meter Figure 3-1. Standard 4540 Series RF Power Meter - Front Panel Table 3-1 Operating Controls, Indicators and Connections Reference # Front Rear Nomenclature Function 1 1 Internal Calibrator The output of the built-in 50MHz programmable calibrator is available from a Type-N connector located on the front, or optionally on the rear panel of the instrument.
Boonton 4540 Series RF Power Meter Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear Nomenclature Function 4 ◄ and ► Keys Used to assist navigating between levels of the menu structure while in Menu Mode or Zero/Cal Mode and to select individual editing numeric parameters. The ◄ key serves as a ―Previous‖ key during menu operation, and will navigate back to the next higher menu in the hierarchy unless the user is in digit editing numeric entry mode.
Boonton 4540 Series RF Power Meter Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear Nomenclature Function 11 Cal/Zero Key Displays the Sensor Calibration menu to select and initiate automatic sensor offset and gain adjustments using the 4540’s built-in 50MHz calibrator or an external calibrator. The menu also permits manual control of the calibrator’s output. Pressing Cal/Zero again will toggle between active channels.
Boonton 4540 Series RF Power Meter Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear Nomenclature Function 20 Cooling Fan Cooling air intake 21 AC Line Input A multi-function power input module is used to house the AC line input, main power switch, and safety fuse. The module accepts a standard AC line cord, included with the power meter. The power switch is used to shut off main instrument power.
Boonton 4540 Series RF Power Meter 3.3 LCD Monitor Display This subsection includes a picture (Figure 3-3) and a table (Table 3-2) of descriptions of the display layout of the 4540 Series. Figure 3-3a shows the graphics (trace) display mode of the instrument. Figure 3-3b shows the pulse mode text (automeasure) display mode of the instrument. Figure 3-3c shows the modulated text mode. In addition there are text formats for reports. See Section 4.5 for more information on the display format.
Boonton 4540 Series RF Power Meter Table 3-2 LCD Monitor Display Fields (continued) Ref. No. Field Name Mode Description 5 Timebase All For pulse and modulated mode, indicates the timebase per division selected for the waveform display. In statistical mode this field is replaced with a label indicating the half span dB level for the horizontal axis. 6 Status message All Displays messages indicating the status of the instrument.
Boonton 4540 Series RF Power Meter Table 3-2 LCD Monitor Display Fields (continued) 10 GPIB/Remote Status All The GPIB/Remote status indicator is located at the lower right corner of the 4540 display. Blank - Remote control is not active and in Local state. REM. Indicates the 4540 is in the Remote state. LSN. Indicates the 4540 is addressed to Listen. . TLK. Indicates the 4540 is addressed to Talk. SRQ. Indicates the 4540 is requesting service from the bus controller.
Boonton 4540 Series RF Power Meter Table 3-2 LCD Monitor Display Fields (continued) Ref. No. Field Name Mode Description 12 Main Measurement Modulated (Text) Displays the main measurement for the channel. This menu selectable field can be channel average power or one of several math functions. 13 Secondary Field Modulated (Text) There are two secondary parameter fields available for each channel’s text display. These fields can display any available measurement or setting for any channel.
Boonton 4540 Series RF Power Meter 3.4 Initialize The procedures presented in this section will initialize the 4540 Series and prepare it for operation. Steps 1 through 3 should be performed every time you turn on the instrument. Step 4 only needs to be performed when you wish to return the instrument operation to a known state. This usually occurs after turning the instrument on or at the beginning of a new test. If you have completed Subsection 2.
Boonton 4540 Series RF Power Meter Table 3-3.
Boonton 4540 Series RF Power Meter Table 3-3. Initialized Parameters (continued) Parameter Applies to: Default Parameters Related to the Time > Menu Timebase Timebase Position Position Control Trig Delay Pulse Mode Modulated Mode Pulse Mode Pulse Mode Pulse Mode 100 µs/Div 200 ms/Div 5.
Boonton 4540 Series RF Power Meter Table 3-3. Initialized Parameters (continued) Parameter Parameters Related to the Stat Mode > Menu Horizontal Scale Horizontal Offset Term Options (Menu) Term Action Term Count Term Time Cursors (Menu) Cursor Mode Cursor Percent Cursor Power Reference Applies to: Default Statistical Mode Statistical Mode 2 dB/Div 0.00 dB Statistical Mode Statistical Mode Statistical Mode Decimate 10 Megasamples 3600 s (not changed by init) Percent 1.0000 % 0.
Boonton 4540 Series RF Power Meter 3.5 Calibration Before any pulse measurements can be acquired with the 4540 Series, a peak power sensor must be connected from the instrument to the built-in or external calibrator, and calibrated. Making measurements with a CW sensor can be accomplished without calibration. However, it is recommended that autocalibration be performed for optimal performance for CW sensors as well. RF voltage probes do not require autocalibration.
Boonton 4540 Series RF Power Meter STEP 1. PROCEDURE Select the Calibration menu soft key. Initiate autocalibration by selecting AutoCal > Start. The AutoCal routine will calibrate the entire dynamic range of the sensor in approximately 1.5 minutes while reporting status via a dialog box. Figure 3-4 Autocalibration screen. Note 2. You can access the calibration menus directly using the Cal/Zero key. Pressing the key repeatedly will cycle between each channel’s calibration menu.
Boonton 4540 Series RF Power Meter Figure 3-5.
Boonton 4540 Series RF Power Meter 3.6 Making Measurements What you need to know To perform accurate measurements, the following is a minimum list of things you should know about the signal that you wish to measure. Signal frequency - The center frequency of the carrier must be known to allow sensor frequency response compensation. Modulation Bandwidth - If the signal is modulated, know the type of modulation and its bandwidth.
Boonton 4540 Series RF Power Meter 3.6.2 Pulse Mode For periodic or pulsed signals, it is often necessary to analyze the power for a portion of the waveform, or a certain region of a pulse or pulse burst. For these applications, the 4540 Series has a triggered Pulse Mode. Operation in this mode is similar to a digital storage oscilloscope - power samples are stored in a circular memory buffer until a trigger signal is received.
Boonton 4540 Series RF Power Meter 3.6.3 Statistical Mode Certain signals are completely random and provide no event that can serve as a trigger for measurements. CDMA or OFDM are common examples. The 4540 Series Statistical Mode was designed to provide measurements for these type of signals. Statistical Mode is only effective when a peak power sensor is connected to the 4540. It is the best choice for analyzing ―noise-like‖ signals that are modulated in a random, non periodic fashion.
Boonton 4540 Series RF Power Meter Note that this analysis does not depend upon any particular test signal, or upon synchronization with the modulating signal. In fact, the analysis can be done using actual communication system signals. Normal operation is not disturbed by the need to inject special test signals.
Boonton 4540 Series RF Power Meter 4. Operation This section presents the control menus and procedures for operating the 4540 Series in the manual mode. All the display menus that control the instrument are illustrated and accompanied by instructions for using each menu item. Note Before any pulse measurements can be acquired with the 4540 Series, a peak power sensor must be connected from the instrument to the built-in or external calibrator, and calibrated.
Boonton 4540 Series RF Power Meter 4.3 Numerical Data Entry Many 4540 Series parameters can be entered directly from the numeric keypad. In some cases, such as the frequency of the measured signal, keypad entry is necessary. In this example, the following menu path is used. Channel > Channel1 > Extensions Pg 2 > Frequency 1. Assume the frequency is initially set to 1.000 GHz. In this case "1.000 GHz" will appear in the menu box under "Frequency". Further, assume the actual measurement frequency is 578.
Boonton 4540 Series RF Power Meter 4.4 Sensor Calibration The Model 4540 Series features a built-in AutoCal process that automatically calibrates the Peak Power sensors for all measurement modes, and also calibrates most CW Power sensors for the modulated measurement mode. An internal programmable 50 MHz calibrator outputs discrete incremental power levels covering the dynamic range of the sensor. Zeroing is part of the AutoCal process and is adjusted for the High and Low video bandwidths where applicable.
Boonton 4540 Series RF Power Meter 4.5 Menu Reference 4.5.1 Main Menu - Pulse and Modulated Modes - Pg 1 (Top Level) The Main Menu is the topmost menu level from which all other menus originate. Main > Channel Description: Opens the Channel Menu for functions specific to CH1 and CH2. For Modulated Mode only, REF1 and REF2 are added. Menu Type: Menu Selection Main > Measure Description: Opens the Measure Menu to select the measurement mode and related functions.
Boonton 4540 Series RF Power Meter 4.5.2 Main Menu - Pulse and Modulated Modes - Pg 2 (Top Level) The Main Menu is the topmost menu level from which all other menus originate. Main > Markers Description: Opens the Marker Menu for functions specific to the operation of vertical cursors Marker 1 and Marker 2. Menu Type: Menu Selection Main > Display Description: Opens the Display Menu for functions specific to the displayed images.
Boonton 4540 Series RF Power Meter 4.5.3 Main Menu - Statistical Mode - Pg 1 (Top Level) The Main Menu is the topmost menu level from which all other menus originate. Main > Channel Description: Opens the Channel Menu for functions specific to CH1, CH2, REF1 and REF2. Menu Type: Menu Selection Main > Measure Description: Opens the Measure Menu to select the measurement mode and related functions.
Boonton 4540 Series RF Power Meter 4.5.4 Main Menu - Statistical Mode - Pg 2 (Top Level) The Main Menu is the topmost menu level from which all other menus originate. Main > Cursors Description: Opens the Cursors Menu for functions specific to the operation of the statistical mode cursors. Menu Type: Menu Selection Main > Display Description: Opens the Display Menu for functions specific to the displayed images.
Boonton 4540 Series RF Power Meter 4.5.5 Measure Menu The Measure menu contains items for controlling measurement acquisition and mode. Main > Measure > Measurement Description: Set or return the state of the data acquisition mode for single or free-run measurements. If Measurement is set to On, the 4540 immediately begins taking measurements (Modulated and Statistical Modes), or arms its trigger and takes a measurement each time a trigger occurs (Pulse Mode).
Boonton 4540 Series RF Power Meter Main > Measure > Clear Description: Clear display traces and all data buffers for CH1 and CH2. Clears averaging filters to empty. Menu Type: Action. SCPI Command: DISPlay:CLEar Main > Measure > Single Sweep ( Pulse Mode only ) Description: Starts a single measurement cycle in Pulse mode. Enough trace sweeps must be triggered to satisfy the channel averaging setting. Menu Type: Action.
Boonton 4540 Series RF Power Meter 4.5.6 Channel Menu Select the specific menu for the desired measurement or reference channel if available. Main > Channel > Channel 1 Description: Select the Channel 1 Menu. Menu Type: Menu Selection. SCPI Command: None Main > Channel > Channel 2 Description: Select the Channel 2 Menu. Channel 2 is available only on the Model 4542. Menu Type: Menu Selection.
Boonton 4540 Series RF Power Meter 4.5.7 Channel 1 | 2 Menu The Channel menu contains items that affect a single measurement channel. The ―channel‖ is a full measurement path, beginning with an external power sensor and sensor cable, and including data acquisition, processing of measurements, and display of the processed information. The Model 4541 single channel power meter has only a Channel 1 menu, while the dual-channel Model 4542 has separate color-coded menus for Channel 1 and Channel 2.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Calibration Description: Select the calibration menu for the specified channel. Menu Type: Menu Selection. Main > Channel > Channel 1 | 2 > Extensions 4-12 Description: Select the extensions menu for the specified channel. Menu Type: Menu Selection.
Boonton 4540 Series RF Power Meter 4.5.8 Channel > Calibration Menu The calibration menu contains menu items to initiate multi-step autocalibration, fixed calibration, and sensor offset zeroing. Additionally, menus are provided to allow selection of the calibration source and a shortcut menu to the selected calibrator control menu. Note: You can access the calibration menus directly using the Cal/Zero key. Pressing the key repeatedly will cycle between each channel’s calibration menu.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Calibration > Sel Calibrator Description: Selects the internal 50 MHz calibrator or the optional 1 GHz external calibrator for autocal and zero. The selection made will apply globally to all selection points in the menu system. If the optional 1 GHz calibrator is not connected, an error message will appear if an attempt is made to select it. Menu Type: Toggle. SCPI Command: None.
Boonton 4540 Series RF Power Meter 4.5.9 Channel > Extensions Pg 1 Menu (Modulated Mode) The extensions menu provides access to more advanced channel options such as trace filtering and channel units. Additionally access to menu page two is provided. This menu is for sensors in Modulated Mode. Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Units Description: Set the channel units. For power sensors, voltage is calculated with reference to the sensor input impedance.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Filter Time Description: Set the current length of the integration filter on the selected channel. Menu Type: Numeric value. Range: 0.010 s to 15.000 s SCPI Command: SENSe[1|2]:FILTer:TIMe Main > Channel > Channel 1 | 2 > Extensions Pg 1 > More 4-16 Description: Opens page 2 of the Channel Extensions Menu.
Boonton 4540 Series RF Power Meter 4.5.10 Channel > Extensions Pg 1 > Corrections Menu (Modulated Mode) The corrections menu contains items for adjusting correction factors for sensors. This menu consists of an offset adjustment, and a frequency cal factor control. Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > dB Offset Description: Set a measurement offset in dB for the selected sensor.
Boonton 4540 Series RF Power Meter Fig. 4-11 Corrections Menu (Modulated) Voltage Probe 4-18 Fig. 4-12 Corrections Menu (Modulated) Peak Power Sensor - No TC Operation Fig.
Boonton 4540 Series RF Power Meter 4.5.11 Channel > Extensions Pg 2 Menu (Modulated Mode) This menu is page two of the channel extensions menu. This menu provides items to set video bandwidth, sensor correction frequency, and peak hold. You can also access the alarms menu. Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Video BW Description: Set the sensor video bandwidth for the selected sensor. The trigger channel bandwidth is not affected. High is the normal setting for most measurements.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF frequency for the current measurement. The appropriate frequency calfactor from the sensor’s EEPROM table will be interpolated and applied automatically. Application of this calfactor cancels out the effect of variations in the flatness of the sensor’s frequency response.
Boonton 4540 Series RF Power Meter 4.5.12 Channel > Extensions Pg 2 > Alarms Menu (Modulated Mode) The Alarms menu provides controls for enabling alarms and adjusting alarm limits (trip points). Controls the power limit alarm operation. When alarm operation is enabled, the ―primary measurement‖(average power in CW or Modulated modes, average power between markers in Pulse mode) is monitored, and compared to preset upper and lower power limits.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit 4-22 Description: Set the lower limit power level for the selected channel. When the measured average power is below the lower limit, a down arrow ▼ will appear on the display above the units on the main text screen in modulated mode, and flag bits are set in the remote mode alarm register. Menu Type: Numeric value. Range: -300.000 to 300.
Boonton 4540 Series RF Power Meter 4.5.13 Channel > Extensions Pg 1 Menu (Pulse Mode) The extensions menu provides access to more advanced channel options such as trace averaging and channel units. Additionally access to the corrections, define pulse, and extensions menu page two are provided. This menu is for Peak Power Sensors in Pulse Mode. Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Units Description: Set the channel units.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Define Pulse Description: Selects the define pulse menu for the selected channel. Menu Type: Menu Select. Main > Channel > Channel 1 | 2 > Extensions Pg 1 > More 4-24 Description: Opens page 2 of the Channel Extensions Menu.
Boonton 4540 Series RF Power Meter 4.5.14 Channel > Extensions Pg 1 > Corrections Menu (Pulse Mode) The corrections menu contains items for adjusting correction factors for sensors. This menu consists of an offset adjustment, frequency cal factor control, and a control for enabling or disabling temperature compensation for peak power sensors. Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > dB Offset Description: Set a measurement offset in dB for the selected sensor.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle Description: Sets the duty cycle in percent for calculated CW pulse power measurements. Valid only for thermal sensors and CW sensors in the square-law region and subject to the accuracy of the duty cycle value. Setting the duty cycle to 100% is equivalent to a CW measurement. Note that this method of measuring pulse power should be used only if a peak power measurement cannot be used.
Boonton 4540 Series RF Power Meter 4.5.15 Channel > Extensions Pg 1 > Define Pulse Menu (Pulse Mode) The define pulse menu contains items for adjusting how automatic pulse measurements are calculated. Measurements such as rise time, fall time, and pulse width are affected by these adjustments. For more information on pulse calculation definitions, see section 6.2 Pulse Definitions. This menu is only valid for peak power sensors in pulse mode.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Define Pulse > Pulse Units Description: Set the units for entering the pulse distal, mesial, and proximal levels. If Volts are selected, the pulse transition levels will be defined as the specified percentage in voltage. If set to WATTS, the levels are defined in percent power.
Boonton 4540 Series RF Power Meter 4.5.16 Channel > Extensions Pg 1 > Define Pulse > Pulse Gate Menu Main Channel > Channel 1 | 2 > Extensions Pg 1 > Pulse Gate > Start Gate Description: Sets the beginning of the pulse measurement region as a percentage of pulse width. Menu Type: Numeric value. Range: Continuous range 0.0 % to 40.
Boonton 4540 Series RF Power Meter 4.5.17 Channel > Extensions Pg 2 Menu (Pulse Mode) This menu is page two of the channel extensions menu. This menu provides items to set video bandwidth, sensor correction frequency, and peak hold. You can also access the alarms menu. Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Video BW Description: Set the sensor video bandwidth for the selected sensor. The trigger channel bandwidth is not affected. High is the normal setting for most measurements.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF frequency for the current measurement. The appropriate frequency calfactor from the sensor’s EEPROM table will be interpolated and applied automatically. Application of this calfactor cancels out the effect of variations in the flatness of the sensor’s frequency response.
Boonton 4540 Series RF Power Meter 4.5.18 Channel > Extensions Pg 2 > Alarms Menu (Pulse Mode) The Alarms menu provides controls for enabling alarms and adjusting alarm limits (trip points). Controls the power limit alarm operation. When alarm operation is enabled, the ―primary measurement‖(average power in CW or Modulated modes, average power between markers in Pulse mode) is monitored, and compared to preset upper and lower power limits.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit Description: Set the lower limit power level for the selected channel. When the measured average power is below the lower limit, a down arrow ▼ will appear on the display above the units on the main text screen in modulated mode, and flag bits are set in the remote mode alarm register. Menu Type: Numeric value. Range: -300.000 to 300.
Boonton 4540 Series RF Power Meter 4.5.19 Channel > Extensions Pg 1 Menu (Statistical Mode) The extensions menu provides access to more advanced channel options such as corrections and channel units. Additionally access to menu page two is provided. This menu is for peak sensors in Statistical Mode. Main > Channel 1 | 2 > Extensions Pg 1 > Units Description: Set the channel units. For power sensors, voltage is calculated with reference to the sensor input impedance.
Boonton 4540 Series RF Power Meter 4.5.20 Channel > Extensions Pg 1 > Corrections Menu (Statistical Mode) The corrections menu contains items for adjusting correction factors for sensors. This menu consists of an offset adjustment, frequency cal factor control, and a control for enabling or disabling temperature compensation for peak power sensors. Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > dB Offset Description: Set a measurement offset in dB for the selected sensor.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle Description: Menu Type: Numeric Value Range: 0.01% to 100.00% SCPI Command: SENSe[1|2]:CORRection:DCYCle Fig. 4-29 Corrections Menu (Statistical) Voltage Probe 4-36 Sets the duty cycle in percent for calculated CW pulse power measurements. Valid only for thermal sensors and CW sensors in the square-law region and subject to the accuracy of the duty cycle value.
Boonton 4540 Series RF Power Meter 4.5.21 Channel > Extensions Pg 2 Menu (Statistical Mode) This menu is page two of the channel extensions menu. This menu provides items to set video bandwidth, sensor correction frequency, and peak hold. You can also access the alarms menu. Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Video BW Description: Set the sensor video bandwidth for the selected sensor. The trigger channel bandwidth is not affected. High is the normal setting for most measurements.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF frequency for the current measurement. The appropriate frequency calfactor from the sensor’s EEPROM table will be interpolated and applied automatically. Application of this calfactor cancels out the effect of variations in the flatness of the sensor’s frequency response.
Boonton 4540 Series RF Power Meter 4.5.22 Channel > Extensions Pg 2 > Alarms Menu (Statistical Mode) The Alarms menu provides controls for enabling alarms and adjusting alarm limits (trip points). Controls the power limit alarm operation. When alarm operation is enabled, the ―primary measurement‖(average power in CW or Modulated modes, average power between markers in Pulse mode) is monitored, and compared to preset upper and lower power limits.
Boonton 4540 Series RF Power Meter Main > Channel > Channel 1 | 2 > Extensions Pg2 > Alarms > Low Limit Description: Set the lower limit power level for the selected channel. When the measured average power is below the lower limit, a down arrow ▼ will appear on the display above the units on the main text screen in modulated mode, and flag bits are set in the remote mode alarm register. Menu Type: Numeric value. Range: -300.000 to 300.
Boonton 4540 Series RF Power Meter 4.5.23 Channel > Ref 1 | 2 Menu The Reference Menu is used to manually or automatically set a power or voltage reference level or to make the current power or voltage level the reference level. With the reference mode enabled, power or voltage values will be expressed relative to the reference rather than absolute. The text display will also show relative values.
Boonton 4540 Series RF Power Meter Main > Channel > Ref 1 | 2 > Set Ref 4-42 Description: Manually set the ratiometric mode reference level for the selected channel. When the reference level is set, the power specified by the argument will become the current reference level, and all power readings will be in dBr, relative to this level. Menu Type: Numeric value. Range: -200.000 to 200.
Boonton 4540 Series RF Power Meter 4.5.24 Trigger Menu (Pulse Mode) The Trigger menu is used to configure trigger settings for pulse measurements. The provided settings control selection of hardware trigger source, trigger polarity, setting the trigger level, and configuring holdoff timing. This menu is only available in Pulse mode. Main > Trigger > Trig Mode Description: Set the trigger mode for synchronizing data acquisition with pulsed signals.
Boonton 4540 Series RF Power Meter Main > Trigger > Trig Slope Description: Set the trigger slope or polarity. When set to +, trigger events will be generated when a signal’s rising edge crosses the trigger level threshold. When – is selected, trigger events are generated on the falling edge of the pulse. Menu Type: Toggle. Choices: +, - SCPI Command: TRIGger:SOURce Main > Trigger > Holdoff 4-44 Description: Set the trigger holdoff time in microseconds.
Boonton 4540 Series RF Power Meter 4.5.25 Time Menu (Modulated and Pulse Mode) The time menu provides for controlling the timebase setting, trigger position, and trigger delay. Main > Time > Timebase Description: Set the timebase in seconds/division. The 4540 has fixed timebase settings in a 1-2-5 sequence, and if the argument does not match one of these settings, it will be forced to the next highest entry. Note: There are separate timebases for the Pulse Mode and the Modulated Mode.
Boonton 4540 Series RF Power Meter Main > Time > Trig Delay Description: Set the trigger delay time in seconds with respect to the trigger. Positive values cause the actual trigger to occur after the trigger condition is met. This places the trigger event to the left of the trigger point on the display, and is useful for viewing events during a pulse, some fixed delay time after the rising edge trigger.
Boonton 4540 Series RF Power Meter 4.5.26 Stat Mode Menu (Statistical Mode) The statistical mode menu provides for controlling the horizontal axis scale of the CCDF display, the length of time or maximum sample count of the CCDF and the action to be taken when the terminal condition is met. Main Pg1 > Stat Mode > Horiz Scale Description: Select the horizontal scale for statistical graphic displays. Menu Type: Numeric value. Range: 0.
Boonton 4540 Series RF Power Meter 4.5.27 Term Options Menu (Statistical Mode) The statistical mode menu provides for controlling the horizontal axis parameters of the CCDF display, the length of time or maximum sample count of the CCDF and the action to be taken when the terminal condition is met. Main Pg1 > Stat Mode > Term Options > Term Action Description: Select the action to take when either the terminal count is reached or the terminal time has elapsed.
Boonton 4540 Series RF Power Meter 4.5.28 Cursors Menu (Statistical Mode) The statistical cursors menu controls the mode and the vertical or horizontal position of the CCDF graphical measurement cursor. This cursor can operate with either the value of the normalized CCDF (percent probability) or the power setting as the indepenedent variable. Whichever axis is chosen as the independent variable, the orthogonal one will become the calculated result of the CCDF automatically.
Boonton 4540 Series RF Power Meter 4.5.29 Markers Menu (Modulated and Pulse Modes) The Markers menu is used to configure measurement markers at specific points on the processed measurement waveform. Automatic measurements may then be used to retrieve measurements at the two markers and in the interval between them. Markers are used in Pulse Mode and Modulated Mode to perform measurements at or between two time offsets relative to the trigger.
Boonton 4540 Series RF Power Meter 4.5.30 Display Menu The Display menu is the entry point for accessing and editing display configuration parameters. Main Pg 2 > Display > Key Beep Description: Set the audible key beep on or off. Menu Type: Toggle. Choices: On, Off SCPI Command: SYSTem:BEEP Main Pg 2 > Display > Graph Header Description: Provides access to the Graph Header menu, which is used to configure the text fields above the graph area. Menu Type: Menu select.
Boonton 4540 Series RF Power Meter 4.5.31 Display > Graph Header Menu The Graph header menu is used to customize the text fields above the graph in graphical display mode. There is a different group of settings for each measurement mode. Main Pg 2 > Display > Graph Header > Num. Rows Description: Set the number of rows of measurement or parameter text above the Graph. Menu Type: Numeric Value.
Boonton 4540 Series RF Power Meter Table 4-1 Field Parameter Selection List Field Parameter Description Pulse Width Rise Time Fall Time Pulse Period Pulse Rep Freq Duty Cycle Off Time Pulse Peak Power Pulse Average Power Overshoot Waveform Average Power Pulse Top Amplitude Pulse Bottom Amplitude Time Skew Edge Delay Channel Frequency Frequency Calfactor Video Bandwidth Channel Averaging Trigger Settings Trigger Delay Vertical Scale Vertical Center dB Offset Marker Average Marker Max Marker Min Marker Ma
Boonton 4540 Series RF Power Meter 4.5.32 Display > Text Mode Menu The Text Mode menu can be used to configure the Modulated text display. The main measurement sources for channel 1 and channel 2 can be set from this menu. There are also menu items that provide access to configurations form the secondary measurement fields. Main Pg 2 > Display > Text Mode > Ch1 Source Description: Set the channel 1 main measurement source for the Modulated mode text display. Menu Type: Pop-up Selection.
Boonton 4540 Series RF Power Meter 4.5.33 Display > Text Mode > Ch1|2 Options Menu The Ch1|2 Options menu provides methods for selecting the secondary measurement/setting fields for the Modulated mode text display. Main Pg 2 > Display > Text Mode > Ch1|2 Options > Sec. Field 1 Description: Selects the measurement or setting to be displayed in the lower left corner of the Modulated mode text display. Menu Type: Pop-up Selection.
Boonton 4540 Series RF Power Meter Main Pg 2 > Display > Text Mode > Ch1|2 Options > Field 2 Chan Description: Select channel associated with the parameter selected by the Sec Field 2 menu. Menu Type: Toggle. Choices: CH 1, CH 2 Main Pg 2 > Display > Text Mode > Ch1|2 Options > Defaults 4-56 Description: Loads the default settings for the Modulated mode text display for the specified channel. Menu Type: Action.
Boonton 4540 Series RF Power Meter 4.5.34 Display > Backlight Menu The Backlight menu provides methods for adjusting backlight and screen saver settings. Main Pg 2 > Display > Backlight > Disp Brt Description: Set the backlight brightness for the LCD display in percent. Zero percent turns the backlight off. Menu Type: Numeric Value.
Boonton 4540 Series RF Power Meter 4.5.35 System Menu The System menu provides access to system configuration, information, and control functions. Main Pg 2 > System > I/O Config Description: Access the I/O and Configuration menu. Menu Type: Menu Select. Main Pg 2 > System > Calibrator Description: Access the Calibrator menu. Menu Type: Menu Select. Main Pg 2 > System > Reports Description: Access the Reports menu. Menu Type: Menu Select.
Boonton 4540 Series RF Power Meter 4.5.36 System > I/O Config Menu The I/O Config menu provides access to set GPIB address and configure Ethernet, and Multi IO port. Main Pg 2 > System > GPIB Address Description: Set the GPIB Address. Press the menu key to obtain the number entry popup. The address entered appears in the menu box. Menu Type: Numeric Value.
Boonton 4540 Series RF Power Meter 4.5.37 System > I/O Config > Ethernet Menu The Ethernet menu provides access to configuration and control functions for the Ethernet interface. Main Pg 2 > System > I/O Config > Ethernet > DHCP/AutoIP Description: Set the state of DHCP/AutoIP system for the Ethernet port. If DHCP/AutoIP is enabled (On), the instrument will attempt to obtain its IP Address, Subnet Mask, and Default Gateway from a DHCP (dynamic host configuration protocol) server on the network.
Boonton 4540 Series RF Power Meter 4.5.38 System > I/O Config > Ethernet > Extensions Menu The Ethernet Extensions menu contains menu items for displaying and editing the IP address, subnet mask, default gateway. There is also a read-only menu for the MAC address. Main Pg 2 > System > I/O Config > Ethernet > Extensions > IP Address Description: Set the Internet Protocol (IP) address of the Ethernet adapter. Note: If DHCP/AutoIP mode is enabled, this menu is read-only. Menu Type: Numeric Value.
Boonton 4540 Series RF Power Meter Main Pg 2 > System > I/O Config > Ethernet > Extensions > MAC Address 4-62 Description: Displays the MAC Address for the Ethernet adapter. Menu Type: Numeric Value, Read Only. Format: Ethernet MAC address in nn:nn:nn:nn:nn:nn format, where ―nn‖ is a two-digit hexadecimal number.
Boonton 4540 Series RF Power Meter 4.5.39 System > I/O Config > Multi IO Menu This menu selects the signal to be sent to the Multi-I/O BNC connector on the rear panel. Main Pg 2 > System > I/O Config > Multi IO > IO Mode (Off) Description: Alternately opens and closes the IO Mode Popup Menu. All selections are made from the popup menu. Use the Up Arrow ▲ and Down Arrow ▼ keys to select the desired output. Press the Enter key to open the appropriate submenu and close the popup menu.
Boonton 4540 Series RF Power Meter 4.5.40 System > I/O Config > Multi IO > IO Mode Status Out Menu This menu selects a TTL compatible status signal to the Multi-I/O BNC connector on the rear panel. Main Pg 2 > System > I/O Config > Multi IO > IO Mode (Status Out) Description: Alternately opens and closes the IO Mode Popup Menu. All selections are made from the popup menu. Use the Up Arrow ▲ and Down Arrow ▼ keys to select the desired output.
Boonton 4540 Series RF Power Meter 4.5.41 System > I/O Config > Multi IO > IO Mode Rec Out Menu This menu selects an analog recorder ouput signal to the Multi-I/O BNC connector on the rear panel. Main Pg 2 > System > I/O Config > Multi IO > IO Mode Rec Out Description: Alternately opens and closes the IO Mode Popup Menu. All selections are made from the popup menu. Use the Up Arrow ▲ and Down Arrow ▼ keys to select the desired output.
Boonton 4540 Series RF Power Meter Main Pg 2 > System > I/O Config > Multi IO > Scaling (Rec Out) Description: Selects the scaling of the recorder signal. Menu type: Multiple Choice Choices: Manual, Auto SCPI Command: OUTPut:[MIO:]RECOrder:SCALing < character data > SCPI Arguments: AUTO, MANUAL Main Pg 2 > System > I/O Config > Multi IO > Rec Limits (Rec Out) Description: Opens the Recorder Limits Menu for manual scaling only. Menu type: Menu Selection 4.5.
Boonton 4540 Series RF Power Meter 4.5.43 System > I/O Config > Multi IO > IO Mode Trig Out Menu This menu selects trigger related signals to the Multi-I/O BNC connector on the rear panel. Main Pg 2 > System > I/O Config > Multi IO > IO Mode Trig Out Description: Alternately opens and closes the IO Mode Popup Menu. All selections are made from the popup menu. Use the Up ▲ and Down ▼ Arrow keys to select the desired output. Press the Enter key to open the appropriate submenu and close the popup menu.
Boonton 4540 Series RF Power Meter 4.5.44 System > I/O Config > Multi IO > IO Mode Volt Out Menu This menu selects a voltage level to the Multi-I/O BNC connector on the rear panel. Main Pg 2 > System > I/O Config > Multi IO > IO Mode (Voltage Out) Description: Alternately opens and closes the IO Mode Popup Menu. All selections are made from the popup menu. Use the Up Arrow ▲ and Down ▼ Arrow keys to select the desired output. Press the Enter key to open the appropriate submenu and close the popup menu.
Boonton 4540 Series RF Power Meter 4.5.45 System > Calibrator Menu (Internal 50 MHz) The Calibrator Menu is used to select and control the internal 50 MHz RF calibrator. It can also select the optional, external 1 GHz accessory calibrator (Model 2530). Note that this menu does not contain any items related to sensor calibration - it is only for controlling the calibrator for use as a signal source. For sensor calibration information, refer to Section 3.5 and Section 4.5.2 of this manual.
Boonton 4540 Series RF Power Meter 4.5.46 System > Calibrator Menu (External 1 GHz - optional) The Calibrator Menu is used to seclect and control the external 1 GHz RF Calibrator (Model 2530). It can also select the 50 MHz internal calibrator. The external calibrator may be pulse modulated using either a built-in pulse generator, or via a rearpanel BNC pulse input.
Boonton 4540 Series RF Power Meter Main Pg 2 > System > Calibrator > Status Description: This menu item is not implemented at this time. SCPI Command: None Main Pg 2 > System > Calibrator > Pulse Description: Selects the pulse menu for the external 1 GHz calibrator. This menu item is specific to the 1 GHz calibrator and does not appear when the internal calibrator is selected. Menu Type: Menu Select. Fig.
Boonton 4540 Series RF Power Meter 4.5.47 System > Calibrator > Pulse - Preset Menu (Ext 1 GHz - opt.) The pulse menu is specific to the 1 GHz calibrator when it is connected. Control of the internal pulse generator is availble in 2 forms. Preset mode provides a limited choice of periods and duty-cycles, but is easy to use. The Variable mode provides a wide choice of periods and pulse widths, but with some restrictions.
Boonton 4540 Series RF Power Meter 4.5.48 System > Calibrator > Pulse - Variable Menu (Ext 1 GHz - opt.) The pulse menu is specific to the 1 GHz calibrator when it is connected. Control of the internal pulse generator is availble in 2 forms. The Variable mode provides a wide choice of periods and pulse widths, but has some restrictions. Preset mode provides a limited choice of periods and duty-cycles, but is easy to use.
Boonton 4540 Series RF Power Meter FFF Fig. 4-62.
Boonton 4540 Series RF Power Meter 4.5.49 System > Reports Menu System Reports are text displays that provide additional information for the user. Main Pg 2 > System > Reports > Configuration The Configuration Report shows the hardware options that are licensed and the control software version that is currently loaded. The Meas. and FPGA version information is an integrity check of the software package.
Boonton 4540 Series RF Power Meter Main Pg 2 > System > Reports > Info The Hardware Info Report shows temperature, power supply voltage and data converter readings at critical points within the instrument. It also reports the CPLD program version and diagnostic status for the internal FPGA circuit. This information is useful for troubleshooting and Boonton Service personnel may ask you to report some or all of these readings in certain situations.
Boonton 4540 Series RF Power Meter Main Pg 2 > System > Reports > Remote Cmds The Remote Commands Report is a list of all 4540 SCPI remote programming commands along with a brief description of each. A more complete description can be found in Section 6 of this manual. The menu selections allow you to jump to the beginning or the end of the entire list or to scan up and down the list in pages. You can also use the Up Arrow ▲ and Down Arrow ▼ keys to step up or down one item at a time. Fig.
Boonton 4540 Series RF Power Meter 4.5.50 System > Servicing Menu This menu accesses various service related functions. Main Pg 2 > System > Servicing > Svc Cal Mode Description: Selects the state of Service Cal Mode. This mode enables a group of service functions. Menu Type: Toggle. Choices: On, Off Note: When Off is selected, the Security and Cycle Relays menu boxes are blanked. SCPI Command: None Main Pg 2 > System > Servicing > Set Clock Description: Opens the Set Clock / Calendar menu.
Boonton 4540 Series RF Power Meter 4.5.51 System > Servicing > Security Menu This menu provides a means to erase sensitive stored information that is incidentally stored in the instrument's memories. The two levels of effectivity are documented on the associated displays. Main Pg 2 > System > Servicing > Security > Sanitize Description: Select the Sanitize level which deletes all user data. Menu Type: Action. SCPI Command: None Fig.
Boonton 4540 Series RF Power Meter Main Pg 2 > System > Servicing > Security > Full Sanitize Description: Select the Full Sanitize level which deletes all user data and all operating software. Consider carefully before executing this procedure as it will render the instrument inoperable until all software is re-installed. Menu Type: Action. SCPI Command: None Fig.
Boonton 4540 Series RF Power Meter 4.5.52 System > Sensor Data Menu This menu provides information about the characteristics and conditions of the sensors plugged into Channel 1 and Channel 2 (in the Model 4542 only). Main Pg 2 > System > Sensor Data > Sensor 1 Info Description: See a typical peak sensor report below. Menu Type: Action. SCPI Command: None Fig.
Boonton 4540 Series RF Power Meter Main Pg 2 > System > Sensor Data > Sensor 2 Info Description: See a typical CW sensor report below. Menu Type: Action. SCPI Command: None Fig. 4-70 Sensor Data Report for Sensor 2 Main Pg 2 > System > Sensor Data > Sensor 2 Details 4-82 Description: Similar to Snsr 2 Info but more detailed with raw data dumps. Used by technicians for troubleshooting purposes. NOT ILLUSTRATED. Menu Type: Action.
Boonton 4540 Series RF Power Meter 4.5.53 Setup Menu This menu provides for system preset with default parameters, pre-defined parameters or user defined parameters and automatic acquisition of triggerable signals. Main Pg 2 > Setup > Defaults Description: Load the default values for all parameters. Menu Type: Action. SCPI Command: SYSTem:PRESet Main Pg 2 > Setup > Preset Description: Opens the Preset Menu. Menu Type: Menu Selection.
Boonton 4540 Series RF Power Meter 4.5.54 Setup > Preset Menu This menu provides presets for widely used communications system signals. Main Pg 2 > Setup > Preset > Select Description: Opens the Select Popup Menu with a list of available presets. The current selection appears in the menu box. Use the Up Arrow ▲ and Down Arrow ▼ keys to locate the desired preset. Press the Enter key to copy the selection into the Select menu box and close the popup menu. Use the Esc key to cancel and close the popup.
Boonton 4540 Series RF Power Meter 4.5.55 Setup > User Presets Menu Main Pg 2 > Setup > User Preset > Preset Name Description: Opens the User Preset Popup Menu with a list of User Names that can be associated with a user defined preset. (User1 thru User25). The selected name will appear in the menu box. Menu Type: Action. SCPI Command: None Main Pg 2 > Setup > Preset > Preset Recall Description: Recalls the User Preset selection named above and loads its parameters into the instrument.
Boonton 4540 Series RF Power Meter Fig. 4-75 User Preset Names Popup Menu 4.6 Cal/Zero Key Menu Pressing thefront panel Cal/Zero key will alternately select the color coded Channel 1 and Channel 2 (if present) Calibration menus. For a complete description refer to Section 4.5.8 Channel > Calibration Menu.
Boonton 4540 Series RF Power Meter 5. Remote Operation 5.1 GPIB Configuration The 4540 GPIB interface is configured using the System > I/O Config > GPIB menu. The primary listen/talk address (MLTA) can be set to any value from 1 to 30 inclusive. The value assigned must be unique to each GPIB device. Secondary address is not implemented. ASCII talk and listen terminators are not required or supported – the 4540 supports the industry standard EOI method of terminating strings send to and from the instrument.
Boonton 4540 Series RF Power Meter 5.4 SCPI Language The 4540 Series instruments may be remotely controlled using commands that follow the industry-standard SCPI programming conventions. The default language is: SYSTem:LANGuage SCPI All of the functions of the 4540 Series are accessible remotely via SCPI commands. 5.4.1 SCPI Structure The SCPI instrument model defines a hierarchical command structure based on ―command nodes‖. Each node may contain commands or names of a next-level command node.
Boonton 4540 Series RF Power Meter 5.4.5 Command Arguments and Queries Many commands require arguments. In this case, the entire command string is sent, followed by the argument. A space is used to separate the command from the argument. For example, ―SENSe:CORRection:DCYCle 25.0‖ sets duty cycle correction to a value of 25.0. Arguments may be numeric, or alphanumeric. If a command requires more than one numeric argument, the arguments must be sent as a comma delimited list.
Boonton 4540 Series RF Power Meter :CALCulate:STATe? Query the state of measurement channel 1 (default chan #) :CALC:STAT? Query the state of channel 1 (short form, default chan #) :CALCulate1:STATe? Query the state of measurement channel 1 (specified chan #) :CALC1:STAT? Query the state of channel 1 (short form, specified chan #) :DISPlay:PULSe:TIMEBASE 0.
Boonton 4540 Series RF Power Meter 5.5 Basic Measurement Information The easiest way to obtain a reading is by use of the MEASure command. This command initiates one complete measurement sequence which includes a default configuration. Examples are: MEAS1:POWER? To return the average power of channel 1, or MEAS1:VOLTAGE? To return the average voltage of channel 1. For finer control over the measurement, individual configuration and function commands should be used.
Boonton 4540 Series RF Power Meter 5.6 SCPI Command Reference This section contains a list of all SCPI remote commands accepted by the 4540 Series. The list is grouped by SCPI subsystem or IEEE488.2 function, and includes a detailed description of each command. 5.6.1 IEEE 488.2 Commands The purpose of IEEE488.2 commands is to provide management and data communication instructions for the system by defining a set of ―*‖ commands (an asterisk followed by a three character code).
Boonton 4540 Series RF Power Meter *ESR? Description: Return the current value of the Standard Event Status Register, then clear the register. This register has bits assigned to a number of possible events or conditions of the instrument. When the event occurs, the corresponding bit is latched. The register value is read using this command. Individual bits may be enabled or disabled for SRQ generation using the ESE mask (see *ESE command).
Boonton 4540 Series RF Power Meter *OPC? Description: This command examines the OPC (Operation Complete) status flag and returns a ―1‖ if all pending operations are complete. If pending operations are not yet complete, it does not return. Syntax: *OPC? Returns: Always returns 1 to indicate operations complete. Otherwise, does not return. Valid Modes: Any Description: Return the status of Channel 1 and Channel 2 followed by a list of installed options.
Boonton 4540 Series RF Power Meter *STB? Description: Return the current value of the Status Byte register. This register has bits assigned to a number of possible events or conditions of the instrument. The register value may be read using this command, or may be used to generate a service request (SRQ) over the GPIB when certain conditions exist. Individual bits may be enabled or disabled for SRQ generation using the SRE mask (see *SRE command).
Boonton 4540 Series RF Power Meter *WAI 5-10 Description: Wait command. This command insures sequential, non-overlapped execution. The 4540 always operates in non-overlapped, sequential mode, therefore this command is accepted as valid, but takes no action.
Boonton 4540 Series RF Power Meter 5.6.2 CALCulate Subsystem The CALCulate group of the command subsystem is used to configure post acquisition data processing. Functions in the CALCulate subsystem are used to configure the measurement mode and control which portions of the acquired measurement data is used and how it is processed to yield a finished measurement. In addition to measurement mode, CALCulate is used to define mathematical operations, measurement units, and limit monitoring.
Boonton 4540 Series RF Power Meter CALCulate:LIMit:UPPer[:POWer] Description: Set or return the upper limit power level for the selected channel. This limit is used for level alarms. When the measured average power is above the upper limit, an up arrow ▲ will appear on the display above the units on the main text screen, and flag bits are set in the alarm register which may be accessed using CALCulate:LIMit:FAIL? query and CALCulate:LIMit:CLEAR commands.
Boonton 4540 Series RF Power Meter CALCulate:MATH Description: Set or return the signal source or sources combined in an arithmetic operation for the displayed reading on the selected channel. Ratiometric displays may be made between two sensors, or between a sensor and a stored reference (see CALCulate:REFerence commands), and sum or difference operations may be performed between two sensors, depending on sensor type.
Boonton 4540 Series RF Power Meter CALCulate:REFerence:COLLect Description: For the selected channel, make the current power or voltage reading the reference level for ratiometric measurements, replacing the previous reference level.
Boonton 4540 Series RF Power Meter CALCulate:STATe Description: Set or return the measurement state of the selected channel. When ON, the channel performs measurements; when OFF, the channel is disabled and no measurements are performed. Syntax: CALCulate[1|2]:STATe Argument: = { 0, 1, OFF, ON } Valid Modes: Any CALCulate:UNITs Description: Set or return units for the selected channel. For power sensors, voltage is calculated with reference to the sensor input impedance.
Boonton 4540 Series RF Power Meter 5.6.3 CALibration Subsystem The CALibration group of commands is used to control automatic zero offset and linearity adjustments to the RF power sensor and the channel to which it is connected. Zero offset adjustment can be performed at any time if no RF signal is applied to the sensor. Linearity calibration requires that the sensor be connected to the instrument’s built-in RF calibrator or a Model 2530 1GHz External Calibrator.
Boonton 4540 Series RF Power Meter CALibration[:INTernal]:AUTOcal Description: Performs a multi-point sensor gain calibration of the selected sensor with the internal 50 MHz calibrator. This procedure calibrates the sensor’s linearity at a number of points across its entire dynamic range. The calibration source defaults to the instrument’s internal calibrator if not specified.
Boonton 4540 Series RF Power Meter 5.6.4 DISPlay Subsystem The DISPlay group of commands is used to control the selection and presentation of textual, graphical and TRACe measurements. DISPlay:BACKlight:BRIGhtness Description: Set or return the backlight brightness for the LCD display in percent. Zero percent turns the backlight off.
Boonton 4540 Series RF Power Meter DISPlay:MODUlated:TIMEBASE Description: Set or return the Modulated Mode timebase in seconds/division. The 4540 has fixed timebase settings, primarily in a 1-2-5 sequence. If the argument does not match one of these settings, it will be forced to the next highest entry. Optional units: minutes.
Boonton 4540 Series RF Power Meter DISPlay:PULSe:TSPAN (Note: This command is supplied for compatibility with the Model 4530 Power Meter) Description: Set or return the horizontal time span of the display in pulse mode. Time/Division. Optional units: minutes.
Boonton 4540 Series RF Power Meter DISPlay:[TEXt:]LIN:RESolution Description: Set or return the display resolution for linear power and voltage readings. The number of significant digits displayed is equal to the argument. This command also sets the resolution of measurements returned in remote mode.
Boonton 4540 Series RF Power Meter DISPlay:TRACe:VCENTer Description: Set or return the power or voltage level of the horizontal centerline of the graph for the specified channel in channel units. If a change in the vertical scale causes the center maximum value to be exceeded, the center will be forced to the maximum value for the new range. Syntax: DISPlay:TRACe[1|2]:VCENTer Argument: = Valid Modes: All -200.00 to +200.
Boonton 4540 Series RF Power Meter 5.6.5 FETCh Queries The FETCh? group of queries is used to return specific measurement data from a measurement cycle that has been INITiated and is complete or free-running. FETCh? performs the data output portion of the measurement. FETCh? does not start a new measurement, so a series of FETCh? queries may be used to return more than one set of processed measurements from a complete set of acquired data.
Boonton 4540 Series RF Power Meter FETCh:ARRay:AMEAsure:TIMe? Description: Returns an array of the current automatic timing measurements performed on a periodic pulse waveform. Measurements performed are: the frequency, period, width, offtime and duty cycle of the pulse waveform, and the risetime and falltime of the edge transitions. For each of the measurements to be performed, the appropriate items to be measured must be visible on the display in GRAPH mode.
Boonton 4540 Series RF Power Meter FETCh:ARRay:MARKer:POWer? Description: Returns an array of the current marker measurements for the specified channel. The array consists of the average, maximum, and minimum power and peak-to-average ratio between the two markers, powers at both markers, and the ratio of the two markers. Note the peak-to-average ratio and marker ratio are returned in dB for log units, and percent for linear units.
Boonton 4540 Series RF Power Meter FETCh:INTERval:AVERage? Description: For the specified channel, return the average power or voltage in the time interval between marker 1 and marker 2. The units will be the same as the specified channel. Syntax: FETCh[1|2]:INTERval:AVERage? Returns: CC, average power or voltage between markers Where CC is the measurement condition code.
Boonton 4540 Series RF Power Meter FETCh:INTERval:MINimum? Description: For the specified channel, return the minimum instantaneous power or voltage in the time interval between marker 1 and marker 2. The units will be the same as the specified channel. Syntax: FETCh[1|2]:INTERval:MINimum? Returns: CC, minimum instantaneous power or voltage between markers Where CC is the measurement condition code.
Boonton 4540 Series RF Power Meter FETCh:MARKer:CURsor:PERcent? Description: Return the CCDF cursor y-axis position in percent with respect to the value set by MARKer:POSItion:POWer (CCDF cursor mode is Power Ref). If CCDF cursor mode is Percent, returns user setting. See MARKer:POSItion:POWer and MARKer:POSItion:PERcent. See also Section 6.4 Statistical Mode Automatic Measurements for more information.
Boonton 4540 Series RF Power Meter FETCh:MARKer:MINimum? Description: For the specified channel and marker, return the minimum power or voltage at the marker. The units will be the same as the specified channel. Syntax: FETCh[1|2]:MARKer[1|2]:MINimum? Returns: CC, minimum power or voltage at marker Where CC is the measurement condition code. Valid Modes: Pulse and Modulated. FETCh:MARKer:RATio? Description: For the specified channel return the ratio of MK1 to MK2.
Boonton 4540 Series RF Power Meter FETCh:TEMPerature:AVERage? Description: Return the averaged internal temperature of the specified peak power sensor. Note: If a nontemperature compensated peak sensor is in use, the 4540 displays a warning message if the temperature has drifted more than 4C from the autocal temperature. Syntax: FETCh[1|2]:TEMPerature:AVERage? Returns: CC, Sensor[1|2] Average Temp in degrees C Where CC is the measurement condition code.
Boonton 4540 Series RF Power Meter 5.6.6 INITiate and ABORt Commands The purpose of the INITiate group of commands is to start and control the process of data acquisition once a measurement has been configured. Depending on settings, the 4540 RF Power Meter may be commanded to begin either a single measurement (INITiate:CONTinuous OFF) which stops when complete, or enter a ―free-run‖ mode where data acquisition occurs continuously (INITiate:CONTinuous ON).
Boonton 4540 Series RF Power Meter 5.6.7 MARKer Subsystem The MARKer group of commands is used to configure and locate measurement markers (cursors) at specific points on the processed measurement waveform. FETCH? or READ? queries may then be used to retrieve measurements at the two markers and in the interval between them.
Boonton 4540 Series RF Power Meter MARKer:POSItion:TIMe Description: Set or return the time (x-axis-position) of the selected marker relative to the trigger. Note that time markers must be positioned within the time limits of the trace window in the graph display. If a time outside of the display limits is entered, the marker will be placed at the first or last time position as appropriate.
Boonton 4540 Series RF Power Meter 5.6.8 MEASure Queries The MEASure group of commands is used to acquire data using a set of high level instructions. They are structured to allow the user to trade off fine control of the measurement process for easy operability. MEASure? provides a complete capability where the power meter is configured, a measurement taken, and results returned in one operation.
Boonton 4540 Series RF Power Meter 5.6.9 MEMory Subsystem The MEMory group of commands is used to save and recall instrument operating configurations, and to edit and review usersupplied frequency dependent offset (FDOF) tables for external devices in the signal path. Up to four configurations may be saved, and two frequency dependent offset tables. Note, however that assigning a stored FDOF table to a particular measurement channel is not a MEMory command; it is handled through the SENSe subsystem.
Boonton 4540 Series RF Power Meter MEMory:SNSR:INFO? Description: Return the sensor ID and parameter data. Syntax: MEMory:SNSR[1|2]:INFO? Argument: None Valid Modes: Any MEMory:SNSR:LFLIN? Description: Return sensor smart shaping table. Syntax: MEMory:SNSR[1|2]:LFLIN? Argument: None Valid Modes: Any MEMory:SNSR:MESSage? Description: Return the sensor message data.
Boonton 4540 Series RF Power Meter MEMory:SYS:LOAD Description: Recall a previously stored file image of an instrument setup. Syntax: MEMory:SYS:LOAD Argument: Alphanumeric filename, "USER1" thru "USER25" Valid Modes: Any MEMory:SYS:STORe Description: Save a file image of the current instrument setup.
Boonton 4540 Series RF Power Meter 5.6.10 OUTPut Subsystem The OUTPut group of commands is used to control various outputs of the 4540. These outputs include the internal 50 MHz calibrator, the multi IO port, and an external 1GHz RF calibration source (Boonton Model 2530). The internal 50 MHz calibrator is primarily used to for automatic calibration of power sensors. Precise level continuous wave (CW) signals can also be sourced by the internal calibrator.
Boonton 4540 Series RF Power Meter OUTPut:EXTernal:PRESent? Description: Return true if the Model 2530 external 1 GHz calibrator is connected and powered up. Otherwise, return false. Syntax: OUTPut:EXTernal:PRESent? Returns: = { 0, 1 }, Accepts { 0, 1, ON, OFF } Valid Modes: Any OUTPut:EXTernal:PULSe:CTRL Description: Set or return the pulse control mode of the external 1 GHz calibrator’s pulse modulator. This parameter has no effect in the External source pulse control mode.
Boonton 4540 Series RF Power Meter OUTPut:EXTernal:PULse:PERWID Description: Set or return the pulse period and pulse width in seconds in the Variable pulse control mode of the external 1 GHz calibrator. This parameter has no effect in the preset pulse control mode or the external pulse source mode. Syntax: OUTPut:EXTernal:PULse:PERWID Argument: = 28e-6 to 131.07e-3 (Pulse period) = 7e-6 to 65.
Boonton 4540 Series RF Power Meter OUTPut:INTernal:SIGNal Description: Set or return the on/off state of the internal 50 MHz output signal. Syntax: OUTPut:INTernal:SIGNal Argument: = { 0, 1, OFF, ON } Valid Modes: Any OUTPut:MIO:MODe Description: Set/return Multi IO mode. Below is a description of each mode: Mode Description Off: Disables multi IO output. Recorder: Analog output driven by 4540 input signal. Voltage: DC voltage output set by user.
Boonton 4540 Series RF Power Meter OUTPut:[MIO:]RECOroderMIN Description: Set or return the Multi IO recorder output minimum, or downscale (-10.0V or 0.0V) power reference level. For voltage probes, this is the equivalent power at the current impedance setting. Syntax: OUTPut:RECOrder:MIN < numeric_value > Argument: < numeric_value > = -200.00 to +200.00 dBm Valid Modes: Any OUTPut:[MIO:]RECOroder:POLarity Description: Set or return the recorder output polarity. UNIPOLAR selects 0.0 volts and 10.
Boonton 4540 Series RF Power Meter OUTPut:MIO:STATus:SETTing Description: Set/return Multi IO status out setting. Below is a description of each argument: Argument Description ALMACT: Sets the Multi IO output to +5 VDC when an alarm is currently active. Multi IO output will be 0 VDC if no alarms are active. ALMLATCH: Sets the Multi IO output to +5 VDC when an alarm limit has been exceeded. Multi IO output will be 0 VDC if no alarm limits have been exceeded.
Boonton 4540 Series RF Power Meter OUTPut:MIO:TRIGout:SOURce Description: Set/return Multi IO source for trigger out mode. Below is a description of each argument: Argument Description INT1, INT2: Multi IO output is active when selected internal trigger is active. EXT: Multi IO output is active when external trigger is active. SWEEP: Pulses when actual trigger event occurs.
Boonton 4540 Series RF Power Meter 5.6.11 READ Queries The purpose of the READ? group of queries is to initiate a measurement cycle, acquire data, and return specific measurement data. READ? performs the initiation, data acquisition, postprocessing, and data output portions of the measurement. READ? is equivalent to ABORting any operation in progress, INITiating a new measurement, then FETChing the data when it is ready. READ? generally does not return data unless acquisition is complete.
Boonton 4540 Series RF Power Meter READ:ARRay:AMEAsure:TIMe? Description: Returns an array of the current automatic timing measurements performed on a periodic pulse waveform. Measurements performed are: the frequency, period, width, offtime and duty cycle of the pulse waveform, and the risetime and falltime of the edge transitions. For each of the measurements to be performed, the appropriate items to be measured must be visible on the display in GRAPH mode.
Boonton 4540 Series RF Power Meter READ:ARRay:MARKer:POWer? Description: Returns an array of the current marker measurements for the specified channel. The array consists of the average, maximum, and minimum power and peak-to-average ratio between the two markers, powers at both markers, and the ratio of the two markers. Note the peak-to-average ratio and marker ratio are returned in dB for log units, and percent for linear units.
Boonton 4540 Series RF Power Meter READ:INTERval:MINFilt? Description: For the specified channel, return the minimum filtered power or voltage in the time interval between marker 1 and marker 2. The units will be the same as the specified channel. Syntax: READ[1|2]:INTERval:MINFilt? Returns: CC, minimum filtered power or voltage between markers Where CC is the measurement condition code.
Boonton 4540 Series RF Power Meter READ:MARKer:AVERage? Description: For the specified channel and marker, return the average power or voltage at the marker. The units are the same as the specified channel. Syntax: READ[1|2]:MARKer[1|2]:AVERage? Returns: CC, average power or voltage at marker Where CC is the measurement condition code. Valid Modes: Pulse and Modulated READ:MARKer:DELTa? Description: For the specified channel return the difference between MK1 and MK2.
Boonton 4540 Series RF Power Meter READ:MARKer:CURsor:PERcent? Description: Return the CCDF cursor y-axis position in percent with respect to the value set by MARKer:POSItion:POWer (CCDF cursor mode is set to Power Ref). If CCDF cursor mode is Percent, returns user setting. See MARKer:POSItion:POWer and MARKer:POSItion:PERcent. See also Section 6.4 Statistical Mode Automatic Measurements for more information.
Boonton 4540 Series RF Power Meter READ:MARKer:RDELTa? Description: For the specified channel return the difference between MK2 and MK1. The units will be the same as marker units. Syntax: READ[1|2]:MARKer:RDELTa? Returns: CC, (MK2-MK1) Where CC is the measurement condition code. Valid Modes: All READ:MARKer:RRATio? Description: For the specified channel return the ratio of MK2 to MK1. The units will be dB for logarithmic units or percent for linear units..
Boonton 4540 Series RF Power Meter 5.6.12 SENSe Subsystem The purpose of the SENSe command subsystem is to directly configure device specific settings used to make measurements, generally parameters related to the RF power sensor and signal processing. The SENSe commands are used to configure the power meter for acquiring data. SENSe enables you to change measurement parameters such as filtering or averaging, sensor bandwidth, operating frequency and calfactors, and measurement gain or offset.
Boonton 4540 Series RF Power Meter SENSe:ARRay:CALTemp? Description: Returns both channels’ sensor calibration temperatures. Syntax: SENSe:ARRay:CALTemp? Returns: CC1, Sensor1 cal temp in degrees C, CC2, Sensor2 cal temp in degrees C Where CCn is the measurement condition code. Valid Modes: All SENSe:CALTemp? Description: Returns the selected channel’s sensor calibration temperature.
Boonton 4540 Series RF Power Meter SENSe:CORRection:FREQuency Description: Set or return the RF frequency for the current sensor, and apply the appropriate frequency calfactor from the sensor’s EEPROM table. Application of this calfactor cancels out the effect of variations in the flatness of the sensor’s frequency response.
Boonton 4540 Series RF Power Meter SENSe:FILTer:STATe Description: Set or return the current setting of the integration filter on the selected channel. OFF provides no filtering, and can be used at high signal levels when absolute minimum settling time is required. ON allows a user-specified integration time, from 2 milliseconds to 15 seconds (see SENSe:FILTer:TIMe command). Note that setting the filter time will force the state to ON.
Boonton 4540 Series RF Power Meter SENSe:PULSe:ENDGT Description: Set or return the point on a pulse, which is used to define the end of the pulse’s ―active‖ interval. This point is defined in percent of the total pulse duration, with 0% corresponding to the midpoint of the rising edge, and 100% corresponding to the midpoint of the falling edge, as defined by the mesial setting.
Boonton 4540 Series RF Power Meter SENSe:PULSe:STARTGT Description: Set or return the point on a pulse, which is used to define the beginning of the pulse’s ―active‖ interval. This point is defined in percent of the total pulse duration, with 0% corresponding to the midpoint of the rising edge, and 100% corresponding to the midpoint of the falling edge, as defined by the mesial setting.
Boonton 4540 Series RF Power Meter 5.6.13 SENSe:MBUF Data Array Commands The MBUF commands (which are grouped under the SENSe subsystem) are used to control buffered data capture and output of a long array of sequential readings (the ―Measurement BUFfer‖). This capability is useful for capturing readings at a rate higher than the host system can read back in real time.
Boonton 4540 Series RF Power Meter SENSe:MBUF:SIZe Description: Set or return the size of the measurement buffer. This controls how many readings will be buffered before the buffer is considered full. If set to zero, the measurement buffer is disabled, and the instrument operates in normal mode. When the buffer is full, no more reading will be placed into the buffer. If INITiate:CONTinuous is OFF, measurements will be considered complete, and the instrument will stop when the buffer has filled.
Boonton 4540 Series RF Power Meter SENSe:MBUF:POSition Description: Return the current acquisition (write) position into the measurement buffer, which is the index of the next buffer point which will be written to. This query is useful for determining whether the measurement buffer is full. A value of zero indicates that no points have been acquired into the buffer yet, and a value equal to SIZE indicates that the buffer is full.
Boonton 4540 Series RF Power Meter SENSe:MBUF:MEASurement Description: Set or return the desired measurement value to be placed into the measurement buffer for the selected channel. In Pulse mode, the value may be chosen to be the average, minimum or maximum value between markers for each sweep event. In Modulated mode, the value may be chosen to be the filtered average, minimum or maximum power reading for each buffer update event. If not specified, this parameter defaults to AVERage.
Boonton 4540 Series RF Power Meter 5.6.14 STATus Commands The STATus command subsystem enables you to control the SCPI defined status reporting structures. The user may examine the status or control status reporting of the power meter by accessing the Device, Operation and Questionable status groups Note that GPIB commands may be used to monitor the instrument’s IEEE-488.2 registers (See GPIB - IEEE-488.2 Commands).
Boonton 4540 Series RF Power Meter STATus:DEVice:EVENt? Description: Returns the current contents of the Device Event register then resets the register value to 0. The Device Event register contains the latched events from the Device Condition register as specified by the Device status group’s positive and negative transition filters.
Boonton 4540 Series RF Power Meter STATus:OPERation:CONDition? Description: Return the current value of the Operation Condition register. The following table shows the bit assignments in the register. These bits are updated by the instrument in real time, and can change in response to changes in the instrument’s operating condition.
Boonton 4540 Series RF Power Meter STATus:OPERation:NTRansition Description: Set or return the value of the negative transition filter bitmask for the Operation status group. Setting a bit in the positive transition filter causes a 1 to 0 (negative) transition in the corresponding bit of the Operation Condition register to cause a 1 to be written in the associated bit of the Operation Event register.
Boonton 4540 Series RF Power Meter STATus:QUEStionable:CONDition? Description: Return the current value of the Questionable Condition register. The following table shows the bit assignments in the register. These bits are updated by the instrument in real time, and can change in response to changes in the instrument’s operating condition.
Boonton 4540 Series RF Power Meter STATus:QUEStionable:EVENt? Description: Returns the current contents of the Questionable Event register then resets the register value to 0. The Questionable Event register contains the latched events from the Questionable Condition register as specified by the Questionable status group’s positive and negative transition filters.
Boonton 4540 Series RF Power Meter STATus:QUEStionable:CALibration:CONDition? Description: Return the current value of the Questionable Calibration Condition register. This register is used to notify the user of questionable quality with respect to calibration. The following table shows the bit assignments in the register. These bits are updated by the instrument in real time, and can change in response to changes in the instrument’s operating condition.
Boonton 4540 Series RF Power Meter STATus:QUEStionable:CALibration:NTRansition Description: Set or return the value of the negative transition filter bit mask for the Questionable Calibration status group. Setting a bit in the positive transition filter causes a 1 to 0 (negative) transition in the corresponding bit of the Questionable Calibration Condition register to cause a 1 to be written in the associated bit of the Questionable Calibration Event register.
Boonton 4540 Series RF Power Meter 5.6.15 SYSTem Subsystem The SYSTem group of commands is used to control system-level functions not directly related to instrument measurement performance. SYSTem commands are used to return error codes or messages from the power meter error queue, control hardware features (backlight and key beep), access the internal clock/calendar, and configure communication parameters for the GPIB and LAN interfaces.
Boonton 4540 Series RF Power Meter SYSTem:COMMunicate:LAN:ADDRess Description: Set or return the IP address for the Ethernet port. Syntax: SYSTem:COMMunicate:LAN:ADDRess Argument: = instrument IP address in nnn.nnn.nnn.nnn (―dot decimal‖) format Valid Modes: Any for queries. DHCP/AutoIP must be disabled (OFF) to set the instrument IP address. SYSTem:COMMunicate:LAN:DGATeway Description: Set or return the default gateway for the Ethernet port.
Boonton 4540 Series RF Power Meter SYSTem:COMMunicate:LAN:MAC? Description: Return the instrument’s Ethernet MAC address. Syntax: SYSTem:COMMunicate: LAN:MAC? Returns: = Ethernet MAC address in nn:nn:nn:nn:nn:nn format, where ―nn‖ is a two-digit hexadecimal number. Valid Modes: Any. SYSTem:COMMunicate:LAN:CURRent:ADDRess? Description: Returns the current IP address for the Ethernet port.
Boonton 4540 Series RF Power Meter SYSTem:DATE:MONTH Description: Set or return the month of the real time clock. Syntax: SYSTem:DATE:MONTH Argument: = 1 through 12 Valid Modes: Any SYSTem:DATE:WEEKday? Description: Return the current day of the week from the real time clock. Syntax: SYSTem:DATE:WEEKday? Returns: = 1 thru 7, where 1 = Sunday … 7 = Saturday.
Boonton 4540 Series RF Power Meter SYSTem:ERRor:CODE? Description: Returns the next queued error code number. Note that errors are stored in a ―first-in-first-out‖ queue, so if more than one error has occurred, repeating this command will report the error codes in the sequence they happened. The action of reading an error removes that error from the queue, so once the most recent error has been read, any more queries will report a code of zero.
Boonton 4540 Series RF Power Meter SYSTem:TIMe:HOUR Description: Set or return the hour of the real time clock. Syntax: SYSTem:TIMe:HOUR Argument: = 0 though 23 Valid Modes: Any SYSTem:TIMe:MINUTE Description: Set or return the minute of the real time clock. On entry, resets seconds to :00.
Boonton 4540 Series RF Power Meter 5.6.16 TRACe Data Array Commands The TRACe group of commands is used to control the output of acquired measurement arrays, which appear as a display trace when the power meter is in Graph mode. The TRACe commands allow outputting a channel’s entire internal display trace (501 measurement points) as one large array, or selecting and returning the array in smaller portions.
Boonton 4540 Series RF Power Meter TRACe:COUNt Description: Set or return the number of trace points, which will be returned each time the TRACe:DATA? query is issued. At the completion of each read, INDEX is automatically incremented by COUNT. If COUNT is set to a number greater than the number of points remaining in the trace, the array will be truncated. Setting COUNT to 501 (and INDEX to zero each time) will return the entire trace array.
Boonton 4540 Series RF Power Meter 5.6.17 TRIGger Subsystem The TRIGger group of commands is used to control synchronization of data acquisition with external events. TRIGger commands generally affect only Pulse Mode. TRIGger:CDF:COUNt Description: Set or return the terminal count (sample population size) in millions of samples for Statistical Mode acquisition.
Boonton 4540 Series RF Power Meter TRIGger:DELay Description: Set or return the trigger delay time in seconds with respect to the trigger for the trigger display location in the LEFT position. Positive values cause the actual trigger to occur after the trigger condition is met. This places the trigger event to the left of the trigger point on the display, and is useful for viewing events during a pulse, some fixed delay time after the rising edge trigger.
Boonton 4540 Series RF Power Meter TRIGger:MODe Description: Set or return the trigger mode for synchronizing data acquisition with pulsed signals. NORM mode will cause a sweep to be triggered each time the power level crosses the preset trigger level in the direction specified by TRIGger:SLOPe. If there are no edges that cross this level, no data acquisition will occur.
Boonton 4540 Series RF Power Meter TRIGger:SOURce Description: Set or return the trigger source used for synchronizing data acquisition. The CH1 and CH2 (Model 4542 only) settings use the signal from the associated sensor. EXT setting uses the signal applied to the rear panel TRIG IN connector.
Boonton 4540 Series RF Power Meter 5.6.18 SCPI Command Summary Table 5-1.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) CALCulate[1|2]:PKHLD CALCulate[1|2]:REFerence:COLLect Set/return state of the peak hold function. = OFF, AVG, INST Set the current reading to be the ratiometric measurement reference level. CALCilate[1|2]:REFerence:DATA Set or return the ratiometric measurement reference level in dBm. = -200.00 to +200.00 dBm CALCulate[1|2]:REFerence:STATe Set or return the ratiometric measurement mode state.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) DISPlay:SCREensaver:STATe Set/return state of screensaver mode. = { 0, 1, OFF, ON } DISPlay:SCREensaver:TIMe Set/return time until display dims. = 1 through 180 DISPlay:[TEXt:]LIN:RESolution Set number of significant digits for linear displays and remote return values. = 3 to 5 DISPlay:[TEXt:]LOG:RESolution Set number of decimal places for log displays and remote return values.
Boonton 4540 Series RF Power Meter 5-1. SCPI COMMAND SUMMARY (continued) FETCh[1|2]:INTERval:PKAVG? Return peak to average power between MK1 and MK2. FETCh:KEY? Return code of last key depressed FETCh[1|2]:MARKer[1|2]:AVERage? Return current reading at the specified marker. FETCh[1|2]:MARKer:DELTa? Return difference between MK2 and MK1.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) MEMory[1|2]:SNSR:CFFAST? Return the sensor high bandwidth (FAST) frequency cal-factor table. MEMory[1|2]:SNSR:CFSLOW? Return the sensor low bandwidth (SLOW) frequency cal-factor table. MEMory[1|2]:SNSR:CWRG Return sensor AC cal data. MEMory[1|2]:SNSR:CWSH? Return sensor smart shaping table. MEMory[1|2]:SNSR:INFO? Return the sensor ID and parameter data. MEMory[1|2]:SNSR:LFLIN? Return sensor smart shaping table.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) OUTPut:EXTernal:SIGNal Set or return the on/off state of the external 1 GHz calibrator output signal. OUTPut:INTernal:LEVel[:POWer] Set or return the power level of the internal 50 MHz calibrator output signal. OUTPut:INTernal:SIGNal Set or return the on/off state of the internal 50 MHz output signal. OUTPut:MIO:MODe Set/return Multi IO mode.
Boonton 4540 Series RF Power Meter Table 5-1.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) SENSe[1|2]:CORRection:CALFactor Set correction factor in dB. = -3.00 to 3.00 SENSe[1|2]:CORRection:DCYCle Set/return duty cycle correction factor in percent. = 0.01 to 100.0% SENSe[1|2]:CORRection:FREQuency Set channel frequency. = 0.001e9 to 110.00e9 Hz SENSe[1|2]:CORRection:OFFSet Set/return sensor offset value in dB.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) SENSe:MBUF:COUNt Set/return number of buffer points to be returned for each query. = 1 to 4096 points SENSe[1|2]:MBUF:INDEX Set/return measurement buffer read index for selected channel. = 0 to 1048575 SENSe[1|2]:MBUF:MEASurement Return measurement buffer array for selected channel. SENSe[1|2]:MBUF:DATA? Set/return measurement buffer read index for selected channel.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) SYSTem:AUTOSET Perform AutoSetup SYSTem:BEEP[:ENABle] Set/return keypad audible beeper status. = 0, 1, OFF, ON SYSTem:BEEP:IMMediate Causes a beep to be emitted. No argument. No return. SYSTem:COMMunicate:GPIB:ADDRess Set or return the GPIB bus address. = 1 to 30 SYSTem:COMMunicate:LAN:ADDRess Set or return the IP address for the Ethernet port. = instrument IP address in nnn.nnn.
Boonton 4540 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) SYSTem:TIMe:MINUTE Set/get minute. = 0 to 59 SYSTem:VERSion? Return SCPI version compliance. = yyyy.v TRACe[1|2]:[MIN|MAX|AVERage]:DATA[:NEXT]? Return count min, max or average power samples starting at index TRACe[1|2]:[MIN|MAX|AVERage]:DUMP? Return entire min, max or average power trace as binary block TRACe[1|2]:COUNt Set/return total number of trace pwr points to return.
Boonton 4540 Series RF Power Meter This page intentionally left blank.
Boonton 4540 Series RF Power Meter 6. Application Notes This section provides supplementary material to enhance your knowledge of the 4540 Series' advanced features and measurement accuracy. Topics covered in this section include pulse measurement fundamentals, automatic measurement principles, and an analysis of measurement accuracy. 6.1 Pulse Measurements 6.1.1 Measurements Fundamentals The following is a brief review of power measurement fundamentals. Unmodulated Carrier Power.
Boonton 4540 Series RF Power Meter Figure 6-1. Pulsed RF Signal Figure 6-2. Distorted Pulse Signal Peak Power. The 4540 Series instruments make power measurements in a manner that overcomes the limitations of the pulse power method and provides both peak power and average power readings for all types of modulated carriers.
Boonton 4540 Series RF Power Meter 6.1.2 Diode Detection Wideband diode detectors are the dominant power sensing device used to measure pulsed RF signals. However, several diode characteristics must be compensated to make meaningful measurements. These include the detector’s nonlinear amplitude response, temperature sensitivity, and frequency response characteristic. Additional potential error sources include detector mismatch, signal harmonics and noise. Detector Response.
Boonton 4540 Series RF Power Meter Frequency Response. The carrier frequency response of a diode detector is determined mostly by the diode junction capacitance and the device lead inductances. Accordingly, the frequency response will vary from detector to detector and cannot be compensated readily. Power measurements must be corrected by constructing a frequency response calibration table for each detector. Mismatch.
Boonton 4540 Series RF Power Meter 6.2 Pulse Definitions IEEE Std 194™-1977 Standard Pulse Terms and Definitions ―provides fundamental definitions for general use in time domain pulse technology‖. Several key terms defined in the standard are reproduced in this subsection, which also defines the terms appearing in the 4540 Series text mode display of automatic measurement results. 6.2.1 Standard IEEE Pulse The key terms defined by the IEEE standard are abstracted and summarized below.
Boonton 4540 Series RF Power Meter Table 6-1. Pulse Terms TERM DEFINITION Base Line The two portions of a pulse waveform which represent the first nominal state from which a pulse departs and to which it ultimately returns. Top Line The portion of a pulse waveform which represents the second nominal state of a pulse. First Transition The major transition of a pulse waveform between the base line and the top line (commonly called the rising edge).
Boonton 4540 Series RF Power Meter 6.3.2 Automatic Measurement Terms The following terms appear in the 4540 Series Text display in the Pulse mode. The Text column lists the abbreviated forms that appear on the display screen. Table 6-2. Automatic Measurement Terms Text TERM DEFINITION Width Pulse Width The interval between the first and second signal crossings of the mesial line.
Boonton 4540 Series RF Power Meter 6.3.3 Automatic Measurement Sequence The automatic measurement process analyzes the captured signal data in the following sequence: 1. Approximately 500 samples of the waveform (equivalent to one screen width) are scanned to determine the maximum and minimum sample amplitudes. 2. The difference between the maximum and minimum sample values is calculated and stored as the Signal Amplitude. 3.
Boonton 4540 Series RF Power Meter 7. The process locates the bottom amplitude (baseline) using the IEEE histogram method. A histogram is generated for all samples in the lowest 12.8 dB range of sample values. The range is subdivided into 64 power levels of 0.2 dB each. The histogram is scanned to locate the power level with the maximum number of crossings. This level is designated the baseline amplitude. If two or more power value have equal counts, the lowest is selected. 8.
Boonton 4540 Series RF Power Meter Item dBm mW Mesial value Sample n Sample n+1 10.0 8.0 11.0 10.0 6.3 12.6 The interpolated crossing time, tx, is calculated from: where P is in watts and n is the number of the sampling interval, referenced to the trigger event. For this example 11. The processor computes the rise and/or falltimes of waveforms that meet the following conditions: a) The waveform must have at least one usable edge (Types 2 through 7).
Boonton 4540 Series RF Power Meter 12. The processor calculates the output values according to the following definitions: a.) Pulse Width Interval between mesial points b.) Risetime See Step 11 c.) Falltime See Step 11 d.) Period Cycle time between mesial points e.) Pulse Repetition Frequency Reciprocal of Period f.) Duty Cycle Pulse Width/Period g.) Off-time (Period) - (Pulse Width) h.) Peak Power Maximum sample value (See Step 1) i.
Boonton 4540 Series RF Power Meter Figure 6-7 Sampling Intervals 14. The processor calculates the delay between the two measurement channels. The time reference for each channel is established by the first signal crossing (starting from the left edge of the screen) which passes through the mesial level(or 50% point in trigger view). The signal excursion must be at least 6 dB in power mode, or 300 mV in trigger-view mode. 6.
Boonton 4540 Series RF Power Meter Fig. 6-9 Statistical Mode Text Display Page 1 Fig. 6-10.
Boonton 4540 Series RF Power Meter Table 6-3. Statistical Automatic Measurements In the Statistical mode the following five automatic measurements are displayed in the 4540 Series Text display for enabled input channels. The Text column lists the abbreviated forms that appear on the display screen. TEXT TERM DEFINITION Avg Average Power: The unweighted avg of all power samples occurring since acquisition started. Peak Peak Power The highest power sample occurring since acquisition was started.
Boonton 4540 Series RF Power Meter The following two global status values are displayed: TEXT DEFINITION Total Time: The total time in Hours:Minutes:Seconds that the data acquisition has been running. Points: The total number of data samples in MSa that has been acquired for each channel in the current run. Note The total number of data samples is affected by the terminal settings.
Boonton 4540 Series RF Power Meter 6.5 Measurement Accuracy The 4540 Series includes a precision, internal, 50 MHz RF reference calibrator that is traceable to the National Institute for Standards and Technology (NIST). When the instrument is maintained according to the factory recommended one year calibration cycle, the calibrator enables you to make highly precise measurements of CW and modulated signals.
Boonton 4540 Series RF Power Meter The worst case approach is a very conservative method in which the extreme conditions of each of the individual uncertainties are added together. If the individual uncertainties are all independent of one another, the probability of all being at their worst-case conditions simultaneously is extremely small. For this reason, the uncertainties are more commonly combined using the RSS method.
Boonton 4540 Series RF Power Meter 6-18 Application Notes
Boonton 4540 Series RF Power Meter The value to use for calibration level uncertainty depends upon the sensor calibration technique used. If AutoCal was performed, the calibrator’s uncertainty at the measurement power level should be used. For sensors calibrated with FixedCal, the calibrator is only used as a single-level source, and you should use the calibrator’s uncertainty at the FixedCal level, (0dBm, for most sensors).
Boonton 4540 Series RF Power Meter Sensor Temperature Coefficient. This term is the error which occurs when the sensor’s temperature has changed significantly from the temperature at which the sensor was AutoCal’d. This condition is detected by the Series 4540 and a ―temperature drift‖ message warns the operator to recalibrate the sensor for drift exceeding ±4C on nontemperature compensated peak sensors.
Boonton 4540 Series RF Power Meter The sensor calfactor uncertainty is due to uncertainties encountered while performing this frequency calibration (due to both standards uncertainty, and measurement uncertainty), and is different for each frequency. Both worst case and RSS uncertainties are provided for the frequency range covered by each sensor, and are listed on the sensor datasheet and in the Boonton Electronics Power Sensor Manual.
Boonton 4540 Series RF Power Meter Step 4: The Source Mismatch Uncertainty is calculated using the formula in the previous section, using the DUT’s specification for ρSRCE and calculating the value ρSNSR from the SWR specification on the 51075’s datasheet. ρSRCE = 0.20 (source reflection coefficient at 10.3GHz) ρSNSR = (1.40 - 1) / (1.40 + 1) = 0.167 (calculate reflection coefficient of 51075, max SWR = 1.40 at 10.3GHz) USourceMismatch = ±2 × ρSRCE × ρSNSR × 100 % = ±2 × 0.20 × 0.167 × 100 % = ±6.
Boonton 4540 Series RF Power Meter Step 10: Now that each of the individual uncertainty terms has been determined, we can combine them to calculate the worstcase and RSS uncertainty values: U (±%) K (U×K)2 ( %2 ) 1. instrument uncertainty 0.10 0.500 0.0025 2. calibrator level uncertainty 2.45 0.577 1.9984 3. calibrator mismatch uncertainty 0.34 0.707 0.0578 4. source mismatch uncertainty 6.68 0.707 22.305 5. sensor shaping error uncertainty 1.00 0.577 0.3333 6. sensor temperature drift uncertainty 0.00 0.
Boonton 4540 Series RF Power Meter Step 2: The Calibrator Level Uncertainty for the Model 2530 External 1GHz calibrator may be calculated from the calibrator’s specification. The 0dBm uncertainty is 0.065dB, or 1.51%. To this figure, we must add 0.03dB or 0.69% per 5dB step from 0dBm. 13dBm is 2.6 5dB steps (13/5) away from 0dBm. Any fraction must always be rounded to the next highest whole number, so we’re 3 steps away. UCalLevel = ±(1.51% + (3 × 0.69%)) = ±3.
Boonton 4540 Series RF Power Meter Step 9: The Sensor Calfactor Uncertainty needs to be interpolated from the uncertainty values in the Boonton Electronics Power Sensor Manual. At 1 GHz, the sensor’s calfactor uncertainty is 2.23%, and at 0.5GHz it is 1.99%. Note, however, that we are performing our AutoCal at a frequency of 1GHz, which is very close to the measurement frequency. This means that the calfactor uncertainty cancels to zero at 1GHz, as discussed in the previous section.
Boonton 4540 Series RF Power Meter 7. Maintenance This section presents procedures for maintaining the 4540 Series. 7.1 Safety Although the 4540 Series has been designed in accordance with international safety standards, general safety precautions must be observed during all phases of operation and maintenance. Failure to comply with the precautions listed in the Safety Summary located in the front of this manual could result in serious injury or death.
Boonton 4540 Series RF Power Meter 7.5 Firmware Upgrade Operating firmware has been loaded into the 4540 Series instrument at the factory. This consists of the proprietary Operating System and the Boonton Electronics 4540 Series Application Firmware. The Application Firmware will be updated from time to time to correct errors and add new features. Users can upgrade their firmware by downloading a special Firmware Upgrade package from the Boonton Electronics webpage, www.boonton.com. 7.
Boonton 4540 Series RF Power Meter How to Start Up in Boot Loader Mode Note 1. From a powered-down state, press and hold the top and bottom blue soft keys simultaneously (located next to the display). 2. Press the green power key. After a few seconds the instrument should display a text screen with the heading ―4540 Bootloader ver. ‖ with a specific version number. 3. The 4540 Peak Power Meter is now ready for firmware upgrade via TCP/IP (LAN).
Boonton 4540 Series RF Power Meter This page intentionally left blank.
Boonton 4540 Series RF Power Meter 8. Appendix A SCPI Error Messages 8.1 SCPI Error Messages NO.
Boonton 4540 Series RF Power Meter NO. MESSAGE -245 "Sensor Disconnected.
Boonton 4540 Series RF Power Meter NO.
Boonton 4540 Series RF Power Meter This page intentionally left blank.
Boonton 4540 Series RF Power Meter 9. Appendix B Model 2530 1GHz Calibrator 9.1 Description The Model 2530 operates as an accessory of the Boonton 4530 or 4540 Series RF Power Meters to provide calibration and test signals at an output frequency of 1.024 GHz. The Model 2530 is controlled by the power meter through its manual and remote programming systems. It appears in the menus for calibration and test signal output and GPIB commands are defined for it.
Boonton 4540 Series RF Power Meter 9.4 Specifications RF Characteristics Operating Modes: Output Frequency: Level Range: Resolution: Source SWR (Refl. Coef.): Accuracy (NIST traceable): Absolute: Linearity: Harmonics: Spurious: CW, Internal or external pulse 1.024 GHz 0.005% -60.0 to +20.0 dBm 0.1 dB 1.20 (0.091) maximum, CW mode +20 to -40 dBm 0.065 dB (1.5%) at 0 dBm 0.
Boonton 4540 Series RF Power Meter 9.5 Connections 1. Plug the supplied RJ-11 cable into the rear panel jack of the 2530 and into the rear panel jack of a BOONTON 4530 or 4540 Series RF Peak Power Meter. Note that the 4540 connection requires use of the enclosed RJ11to-HDB15 adapter (p/n 95125001A) to connect the calibrator cable via the EXT CAL/VGA OUT connector. 2. Connect the AC power cord to a suitable AC power source. 3.
Boonton 4540 Series RF Power Meter This page intentionally left blank.
Boonton 4540 Series RF Power Meter 10. Appendix C Boonton EULA "Boonton Electronics 4540 Series Operating Software" END-USER LICENSE AGREEMENT IMPORTANT-READ CAREFULLY: This End-User License Agreement ("EULA") is a legal agreement between you (either an individual or a single entity) and Boonton Electronics, a subsidiary of the Wireless Telecom Group, Inc.
Boonton 4540 Series RF Power Meter 8. U.S. GOVERNMENT LICENSE RIGHTS. All Product provided to the U.S. Government pursuant to solicitations issued on or after December 1, 1995 is provided with the commercial license rights and restrictions described elsewhere herein. All Product provided to the U.S. Government pursuant to solicitations issued prior to December 1, 1995 is provided with "Restricted Rights" as provided for in FAR, 48 CFR 52.22714 (JUNE 1987) or DFAR, 48 CFR 252.
Boonton 4540 Series RF Power Meter 11. Appendix D Warranty & Repair Repair Policy 4540 Series Instrument. If the Boonton 4540 Series RF Power Meter is not operating correctly and requires service, contact the Boonton Electronics Service Department for return authorization. You will be provided with an RMA number and shipping instructions. Customers outside the USA should contact the authorized Boonton distributor for your area. The entire instrument must be returned in its original packing container.
Boonton 4540 Series RF Power Meter THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUD-ING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PAR-TICULAR PURPOSE. Boonton will not be liable for any incidental damages or for any consequential damages, as defined in Section 2-715 of the Uniform Commercial Code, in connection with products covered by the foregoing warranties.