User Manual

Boonton 4540 Series RF Power Meter
Operation
4-34
4.5.19 Channel > Extensions Pg 1 Menu (Statistical Mode)
The extensions menu provides access to more advanced channel options such as corrections and channel units. Additionally
access to menu page two is provided. This menu is for peak sensors in Statistical Mode.
Main > Channel 1 | 2 > Extensions Pg 1 > Units
Description: Set the channel units. For power sensors, voltage is
calculated with reference to the sensor input impedance.
Note that for ratiometric results, logarithmic units will
always return dBr (dB relative) while linear units return
percent.
Menu Type: Pop-up Selection.
Choices: dBm, Watts, Volts, dBV, dBmV, dBµV
SCPI Command: CALCulate[1|2]:UNIT <character data>
Main > Channel 1 | 2 > Extensions Pg 1 > Corrections
Description: Selects the channel corrections menu.
Menu Type: Menu Select.
Main > Channel 1 | 2 > Extensions Pg 1 > More
Description: Selects Page 2 of the Channel Extensions menu.
Menu Type: Menu Select.
Fig. 4-27 Extensions Pg1
Menu (Statistical)