User manual

84 en | Appendix NurseCall Main Unit
953.92 | v1.0 | 2007.07 User Manual TeleAlarm SA®
Change Day / Night
If you have activated the change Day / Night, the NurseCall system transfers all Alarms during
Night to the group 24. At Day, all groups 00 - 24 can be used. When switching from Day to Night
or vice versa, the Message "DAY-NIGHT" is transmitted to the activated pagers to signalize the
change.
Priority Alarms
The Assistance and Fire Alarms are priority calls transmitted to all activated pagers.
Call repetition
If Alarms/Messages are not acknowledged after about 7 minutes, call repetitions are also sent
to the group 23.
22 977
23 976
24 975
Group User number
Table 13.9 Attribution table