13.6 DECT phone system specifications
13.6.1 Multitone DECT system CS600 and P318 interface
Only Alarms/Messages arriving to the Alarm Buffer are transmitted (except "RELAY OFF"). The
transmission is fully alpha-numeric. The Alarms/Messages are repeated every 1 or 3 minutes
until Acknowledgement (see Section 7.4.13 RS232 message setting (toggle), page 55).
The transmission to the DECT system is a team call. The team number must match with the
paging group programmed for every single
(default paging group= 00).
Change Day / Night
If you have activated the change Day / Night, the NurseCall system transfers all Alarms during
Night to the group 24. At Day, all groups 00 - 24 can be used. When switching from Day to Night
or vice versa, the Message "DAY-NIGHT" is transmitted to the activated pagers to signalize the
change.
Priority Alarms
The Assistance and Fire Alarms are priority calls transmitted to all activated DECT handsets.
Call repetition
If Alarms/Messages are not acknowledged after about 7 minutes, call repetitions are also sent
to the group 23.
i
NOTICE!
The system can transfer the Alarms received to DECT handsets of the types Multitone CH60 or
CH70.
i
NOTICE!
Data rate: 9600 Bauds. Transmission: asynchronous with a 10 bit-structure
(1 startbit, 7 databits with even parity, 1 stopbit) in half-duplex mode.