NurseCall N46 Wall Transmitter en User Manual
NurseCall N46 Wall Transmitter | en 3 Table of Contents 1 Identification 5 1.1 Document 5 1.2 Customer support addresses 6 2 Generalities 7 2.1 NurseCall System 7 2.2 NurseCall N46 Wall Transmitter 7 3 Safety Instructions 8 3.1 Introduction 8 3.1.1 Principle 8 3.1.2 Importance of safety instructions 8 3.1.3 Disregarding safety rules 8 3.2 Environmental conditions 8 3.3 General safety instructions 9 3.3.1 Observation and information 10 3.
en | NurseCall N46 Wall Transmitter 5.6 Wired Inputs 18 5.7 General configuration 19 6 Use 20 6.1 Call for Help 20 6.2 Call for Assistance 20 6.3 Call Acknowledgement (Staff Presence) 21 6.4 Call for Help 2 (reserve) 21 6.5 Technical Alarm 21 6.6 Caregiver Arrival 21 6.7 Caregiver Departure 22 6.8 Daily Message and Battery Low Message 22 6.9 Error Message 22 7 Operation 23 7.1 Description of the six different states 23 7.2 Legend of illustrations 24 7.
NurseCall N46 Wall Transmitter 1 Identification 1.1 Document Identification | en Name No. User Manual 970.030 5 Table 1.1 Document No. Version Description v1.0 | 2009.06 First Edition Table 1.2 Version Management Bosch Security Systems User Manual 970.030 | v1.0 | 2009.
en | Identification 1.
NurseCall N46 Wall Transmitter 2 Generalities 2.1 NurseCall System Generalities | en 7 Alarms and Messages arriving from NurseCall Transmitters such as the NurseCall N46 Wall Transmitter are managed and stored by the NurseCall Main Unit. This unit is compatible with the NurseCall Master. i 2.2 NOTICE! The document "NurseCall General Overview" explains the system concept.
en | Safety Instructions 3 NurseCall N46 Wall Transmitter Safety Instructions ! WARNING! The User / Installer should read and understand this chapter before any intervention on the NurseCall N46 Wall Transmitter. 3.1 Introduction 3.1.1 Principle i 3.1.2 NOTICE! In case of unclear information, please contact your local representative.
NurseCall N46 Wall Transmitter X ! 3.3 Safety Instructions | en 9 Install the NurseCall N46 Wall Transmitter in a dry place, away from any source of heat. CAUTION! Interferences Avoid immediate proximity to other electric devices such as a television. General safety instructions DANGER! ! ! Electrocution During maintenance operations, when the NurseCall N46 Wall Transmitter is powered and its casing is removed, the NurseCall N46 Wall Transmitter may not be left unattended.
en | Safety Instructions ! NurseCall N46 Wall Transmitter CAUTION! Never let any liquid enter the system. In case of liquid spill inside the NurseCall N46 Wall Transmitter, act immediately as follows: 1. Switch off the NurseCall N46 Wall Transmitter by taking out the battery or removing the power supply. 2. i 3.3.1 Dry up the NurseCall N46 Wall Transmitter. 3. Clean the NurseCall N46 Wall Transmitter. 4. Check its electrical functions.
NurseCall N46 Wall Transmitter Description | en 4 Description 4.1 General description 4.1.1 Top view Fig. 4.1 Bosch Security Systems 11 Top view 1. Red emergency call button 2. LED indicator for alarm 3. LED indicator for presence 4. Green acknowledgment button, to acknowledge an alarm locally. 5. Blue assistance call button 6. Socket for coded key insertion (option) 7. Installation compartment 8. Jack plug for pear-push button User Manual 970.030 | v1.0 | 2009.
en | Description 4.2 NurseCall N46 Wall Transmitter Main functions The main function of the NurseCall N46 Wall Transmitter is to enable easier handling of emergencies and help, directly from the patient’s room. 4.2.1 Red emergency call button Its large red button with textured surface makes it easy to identify and to press, launching an emergency call. 4.2.
NurseCall N46 Wall Transmitter Installation | en 5 Installation 5.1 Unpacking 13 The NurseCall N46 Wall Transmitter is carefully packed for transportation. The components contained in the box are protected, but should be handled with care. Store the packaging material for further use (storage or transport). i NOTICE! In case of defective or missing equipment, do not try to install the NurseCall N46 Wall Transmitter. Contact immediately your local representative. 1.
en | Installation NurseCall N46 Wall Transmitter 5.2 Installation 5.2.1 Generalities X ! 5.2.2 Install the NurseCall N46 Wall Transmitter in a dry place, away from any source of heat. CAUTION! Interferences Avoid immediate proximity to other electric devices such as a television. Wall installation The NurseCall N46 Wall Transmitter can be fixed with two to five screws on a wall. Two holes at 60 mm interval are provided for direct mounting on connection boxes.
NurseCall N46 Wall Transmitter 5.2.3 Installation | en 15 Connecting to the power supply The NurseCall N46 Wall Transmitter can be powered by a lithium battery or/and an external 12-24 VDC power supply. If the device is powered with an external power supply, the battery is not necessary, but it can be used as a backup power supply in case of power loss. If the device is powered by the battery only, the battery life is about 2 years. However, it is recommended to change the battery once a year.
en | Installation 5.2.7 NurseCall N46 Wall Transmitter Replacing the battery WARNING! – May explode if exposed to fire; – Use only original batteries intended for use with your NurseCall N46 Wall Transmitter. – Do not expose the battery to liquids; – Do not let the metal contacts on the battery touch another metal.
NurseCall N46 Wall Transmitter 5.2.
en | Installation NurseCall N46 Wall Transmitter 5.3 Inputs and Outputs 5.4 General inputs The * marks a factory setting. External trigger J4 Power supply 5.5 J5 Jack 3.
NurseCall N46 Wall Transmitter 5.7 Installation | en 19 General configuration The * marks a factory setting. Bridge State Function B3 Technical Alarms are not repeated. Open* Soldered Technical Alarms are repeated until alarm acknowledgement (max 20 min.) B4 Open* Calls (except technical alarms) are repeated until call acknowledgement (max 20 minutes) B5 B6 Soldered Calls (except technical alarms) are not repeated. Open* A daily message is sent every 24 hours.
en | Use 6 NurseCall N46 Wall Transmitter Use Every time a NurseCall N46 Wall Transmitter is used by pressing a button or activating a wired input, the device sends a radio signal and the lamps indicates its actual state. Optionally, two relays allow the transmission of status to a remote display such as the signal lamps placed on top of the door. Each radio message is sent with a different code which allows the receiver (NurseCall system) to recognize the performed action. i 6.
NurseCall N46 Wall Transmitter 6.3 Use | en 21 Call Acknowledgement (Staff Presence) The call acknowledgement is triggered by pushing the Acknowledgement button (green). The call acknowledgement stops immediately all repetitions of the calls. Optionally, a call acknowledgement triggers the Staff Presence signal: – by green flashes on the signal lamp and by activating the Staff Presence relay when the Caregiver pushes the Acknowledgement button when entering the room.
en | Use NurseCall N46 Wall Transmitter Insertion of the coded key is signaled by green flashes on the signal lamp and by activating the Staff Presence relay. 6.7 Caregiver Departure The Caregiver Departure message is triggered when the Caregiver removes the coded key from the housing, when leaving the room. The Caregiver Departure message is not repeated. Removing the coded key is signaled by switching off the green flashes on the signal lamp and by releasing the Staff Presence relay. 6.
NurseCall N46 Wall Transmitter Operation | en 7 Operation 7.1 Description of the six different states 23 During operation, the NurseCall N46 Wall Transmitter can be set to one of six following states: State Description Stand-by No call is being processed. Alarm without staff presence The patient has triggered a Call for Help but the Care Personnel has not yet responded. Alarm with staff presence The Care Personnel has responded after a Call for Help.
en | Operation 7.
NurseCall N46 Wall Transmitter Operation | en 25 State before action: Alarm without staff presence Action State Relay contact Signal lamp Radio Message Alarm without staff presence (none) Call for Help Alarm without staff presence (unchanged) Call for Help Call Acknowledgement Alarm with staff presence Acknowledgement Call for Assistance Assistance without staff presence ** Call for Assistance **If the function Assistance only in case of Staff Presence is activated and the NurseCall N46 W
en | Operation NurseCall N46 Wall Transmitter State before action: Assistance with staff presence Action State Relay contact Signal lamp Radio Message Assistance with staff presence (none) Call for Help Assistance with staff presence (unchanged) Call for Assistance Call Acknowledgement Alarm with staff presence Acknowledgement Call for Assistance Assistance with staff presence Call for Assistance State before action: Assistance without staff presence Action State Relay contact Signa
NurseCall N46 Wall Transmitter 7.3.2 Operation | en 27 Device behavior during coded key insertion and removal i NOTICE! The coded key insertion an removal functions are optional.
en | Operation NurseCall N46 Wall Transmitter State before action: Assistance with staff presence Action State Relay contact Signal lamp Assistance with staff presence (none) Assistance with staff Coded key insertion presence (unchanged) Coded key removal Radio Message Staff arrival with code Staff departure with code Assistance without staff presence State before action: Assistance without staff presence Action State Relay contact Signal lamp Radio Message Assistance without staff pr
NurseCall N46 Wall Transmitter Operation | en 7.4 Operation with the staff presence processing deactivated 7.4.
en | Operation 7.4.2 NurseCall N46 Wall Transmitter Device behavior during coded key insertion and removal i NOTICE! The coded key insertion an removal functions are optional.
NurseCall N46 Wall Transmitter 8 Storage | en 31 Storage One can distinguish between long term storage and short term storage. If the NurseCall N46 Wall Transmitter must be temporarily stored, it may not be necessary to wrap it completely. However, if you wish to store the NurseCall N46 Wall Transmitter for a longer time, for instance in a storage room or any similar location, it is recommended to use original packing material. 8.
en | Disposal 9 NurseCall N46 Wall Transmitter Disposal This chapter describes the appropriate disposal of the NurseCall N46 Wall Transmitter. The NurseCall N46 Wall Transmitter is marked with a crossed-out wastebasket symbol. This means that, at the end of its useful lifespan, the product shall be disposed separately from ordinary household wastes in accordance to the EU Directive 2002/96/EC.
NurseCall N46 Wall Transmitter Appendix | en 10 Appendix 10.1 Electrical specifications External power supply 12-24 VDC & consumption Transmission: < 40 mA 33 Idle mode: < 10 mA Battery power supply 3V lithium battery CR2450 & consumption Transmission: < 30 mA Idle mode: < 10 µA Battery life Approx. 2 years Table 10.1 Electrical specifications 10.
en | Appendix 10.4 NurseCall N46 Wall Transmitter EC-Declaration of conformity Fig. 10.1 970.030 | v1.0 | 2009.
TeleAlarm SA Bosch Group Rue du Pont 23 CH-2300 La Chaux-de-Fonds Switzerland Phone +41 32 327 25 40 Fax +41 32 327 25 41 ch.securitysystems@bosch.com www.telealarm.com Bosch Security Systems France Atlantic 361 361, avenue du Général de Gaulle F-92147 Clamart France Phone + 33 (0)825 12 8000 (0, 15 € TTC/Min) Fax + 33 (0)820 900 960 (0, 12 € TTC/Min) fr.securitysystems@bosch.com www.boschsecurity.