User's Manual

Instructions for the Physician
Precision Spectra™ System Information for Prescribers
90668515-01 Rev A 11 of 15
Instructions for the Physician
Implanted Stimulation Devices. If such implanted devices are indicated for the patient,
careful screening is required to determine if safe results can be achieved before permanently
implementing concurrent electrical therapies.
Postural Changes. Depending on the activity level of the patient, postural changes may affect
stimulation intensity. Instruct patients to keep the Remote Control on hand at all times, and
ensure that they understand how to adjust stimulation levels. Refer to Postural Changes in the
Instructions for Patients section of this manual, page 7, for additional information.
Medical Devices/Therapies. If the patient is required to undergo lithotripsy, electrocautery,
external de brillation, radiation therapy, ultrasonic scanning, or high-output ultrasound:
Turn off stimulation at least ve minutes before the procedure or application.
All equipment, including ground plates and paddles, must
be used as far away from the IPG as possible.
Every effort should be taken to keep elds, including current, radiation,
or high-output ultrasonic beams, away from the IPG.
Equipment should be set to the lowest energy setting clinically indicated.
Instruct patients to con rm IPG functionality following treatment by turning
on the IPG and gradually increasing stimulation to the desired level.