Sterilization and Handling
Sterilization
The Advanced Bionics Lead Model SC-2108 and Lead Extension
Model SC-3108 and accessories (except for the Lead Position Labels)
were sterilized with ethylene oxide prior to shipment. Red lines on the
green tape located near the bottom of the inner tray cover indicate
exposure to the sterilization process.
Inspect the condition of the sterilization indicator and the sterile pack-
age before opening the package and using the contents. Do not use the
contents if the indicator lines are not red, if the package is broken or
torn, or if contamination is suspected because of a defective sterile
package seal.
Do not use any component that shows signs of damage.
Do not resterilize the package or the contents. Obtain a
sterile package from Advanced Bionics.
Do not use if “Use Before” date has expired.
Note: The lead, lead extension and accessories are intended for
single use only.
Handling
The lead is designed to perform in the hostile environment of the
human body. Care must be taken to avoid damaging the lead with
sharp instruments or excessive force during surgery. The following
guidelines will help to ensure the longevity of components:
Do not sharply bend or kink the lead or extension.
Do not tie suture(s) directly to the lead or extension
body; use the provided suture sleeves.
Avoid forcing the lead into the epidural space by care-
fully clearing a path using the lead blank.