Physician Lead Manual
Guidelines for Trial-phase Implantation
This section details the recommended procedures for trial-phase tem-
porary implantation of the lead.
Pre-op Instructions
Check that the sterile package is intact. (See “Steriliza-
tion” on page 18.)
Ensure that a Trial Stimulator and Patient Trial Kit are
available for use following lead placement. Install a new
6 volt battery (included in the Patient Trial Kit) in the
Trial Stimulator.
Be sure the Trial Stimulator and Remote Control stimu-
lation settings have been reset. Refer to the IPG manual
for links and resets.
If monopolar testing is anticipated, place the monopolar/
indifferent electrode (available separately) on the
patient’s shoulder or leg and run the cable to the Trial
Stimulator testing site before the patient is prepped and
draped.