Guidelines for Trial-phase Implantation
23
Connecting the OR Cable Assembly
The OR cable extension is designed for temporary connection to the
OR cable to facilitate stimulation testing outside of the sterile field.
After stimulation testing, the cable extension is typically removed and
the OR cable is connected directly to the Trial Stimulator for use dur-
ing the trial phase.
CAUTION: Do not immerse the OR cable connector or plug
in water or other liquids. The OR Cable Assembly is intended
for one-time only use; do not resterilize.
1. If two leads are being implanted, wrap the non-sterile 1-L and 2-
R labels around the cables at the Trial Stimulator to identify lead
connections.
2. Verify that the Trial Stimulator is off.
CAUTION: Always turn the Trial Stimulator off before
connecting or disconnecting the Cable Assemblies.
3. Check that the locking lever on the OR cable connector is in the
open position (0).
4. Slide the proximal end of the lead, with stylet, into the open port
on the OR cable connector.