User Manual

26
Physician Lead Manual
Intraoperative Stimulation Testing
Note: The following steps are for procedural reference only. Please
refer to the BionicNavigator Software Guide for detailed
stimulation testing procedures and guidelines.
1. Test various electrode configurations to obtain paresthesia
.
Note: If lead repositioning is necessary, turn stimulation off before
proceeding.
2. When the desired paresthesia is achieved:
turn the Trial Stimulator off
unlock each OR cable connector and disconnect from
the lead(s)
slowly withdraw the stylet(s)
3. Record the lead position by capturing a fluoroscopic image to be
sure the leads have not moved. Retest if necessary. The image
can also be used for a position comparison at closure to ensure
that the leads did not move.
OPTION A: Temporary Lead Trial
1. Hold the lead distal to the needle hub to maintain lead position
during needle removal.
2. Carefully withdraw the insertion needle from the epidural space
by slowly pulling the needle up towards the proximal end of the
lead.
3. Continue to pull the needle back approximately one centimeter at
a time until the needle tip is exposed.