Guidelines for Trial-phase Implantation
29
5. Once the needle tip is exposed, hold the lead as close to the tip as
possible, then carefully pull the needle completely from the lead.
Anchoring the Lead
1. Place a suture through the supraspinous ligament or deep fascial
tissue.
2. Slide a suture sleeve over the lead and down to the supraspinous
ligament.
3. Ligate the sleeve onto the lead by tying a 2-0 silk or other nonab-
sorbable suture around the center groove of the sleeve to prevent
sliding.
CAUTION: Do not use polypropylene sutures as they may
damage the suture sleeve. Do not suture directly onto the
lead or use a hemostat on the lead body. This may damage
the lead insulation.