User Manual

34
Physician Lead Manual
9. Using the torque wrench supplied, turn the extension connector
setscrew clockwise until it clicks, indicating lock.
Note: Ensure that the wrench is fully seated in the setscrew before
tightening.
The wrench is torque-limited and cannot be overtightened.
10. Form an appropriately-sized pocket using blunt dissection on
either side of midline for coiled excess lead and extension con-
nectors.
11. Place a small loop at the lead for slack. If necessary, loosely tie a
suture around the lead-loop, but do not tighten onto the lead.