User Manual

Patient Remote Control
27
For Rate and Pulse Width: When the desired option screen is dis-
played, first use the
SEL 1 button to find the desired area, then use the
T or S button to decrease or increase the parameter setting for that
area.
To restore a program: Press
SEL 1 to highlight the desired program,
then press
SEL 2 [RESTR]. The restore function returns a program to
the last clinic programmed settings.
Physician Mode
Patient-restricted physician screens provide access to these advanced
programming and interrogation functions:
Communication with the clinician programmer
Electrode and stimulation parameter programming
Electrode impedance monitoring
Remote Control and trial stimulator resetting/re-linking
Entering Physician Mode
1. To access physicians screens, press
SEL 1 and SEL 2 simulta-
neously.
2. The Remote Control will request a code. (To obtain the code,
refer to the Physician Systems Handbook.) Use the S and T
buttons to scroll through the characters, and the
SEL 1 [SEL] but-
ton to advance to the next entry. After entering the code, press
SEL 2 [ENTER].
Once the code is entered, the screen will appear as shown.