User's Manual

Implantable Pulse Generator
Implantable Pulse Generator
90655621-03 Rev A 6 of 11
5. Make a small incision at the desired exit site.
6. Create a subcutaneous tunnel between the lead(s) incision and the IPG pocket site until the
straw is visible and accessible at the exit point.
7. Unscrew and remove the tunneling tool handle.
8. Grasp the tip of the tool with one hand while holding the straw in place with the other hand.
Pull the tunneling tool shaft out through the straw.
9. Push the lead or extension through the straw, then withdraw the straw.
10. Pull the proximal end(s) out of the exit point.
11. Wipe the proximal end(s) clean.