Electromagnetic Emissions and Immunity
Table 3. Guidance and manufacturer’s declaration—electromagnetic
immunity—for equipment and systems that are not life-supporting
a b c
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment—guidance
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Model 6476
LATITUDE Communicator,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended
separation distance
Conducted
RF (IEC
61000-4-6)
3 Vrms
150 kHz to
80 MHz
3 Vrms d = (0.35) ʚP
Radiated
RF (IEC
61000-4-3)
3 V/m
80 MHz to
2.5 GHz
3 V/m d = (3.17) ʚP
(80 MHz to 800 MHz)
d = 6.34 ʚP
(800 MHz to 2.5 GHz)
where P is the maximum
output power rating of