User's Manual

Table Of Contents
41
QTECH™ PROGRAMMER: OPERATION
Figure 33: Setting the desired rst shock energy for debrillation testing
5. Select the Continue button to display the next Induction Test screen.
Note: Ensure that noise markers (“N”) are not present on the S-ECG prior to
induction. The presence of noise markers may delay detection and
therapy delivery.
6. Select and hold the Hold To Induce button for the desired duration (Figure 34).
Figure 34:
Induction Test screen