User's Manual

Table Of Contents
44
QTECH™ PROGRAMMER: OPERATION
Figure 35: Device Status screen
The Device Status overview reports:
Date of the last follow-up session
Total number of shocks delivered since the last follow-up session
Total number of treated episodes since the last follow-up session
Note: Choosing the “View” button in the Treated Episodes row allows navigation
directly to the list of stored episodes (Figure 36).
Total number of untreated episodes since the last follow-up session
Remaining SQ-RX device battery life
Viewing Stored Episodes
The SQ-RX device stores up to 25 treated and 20 untreated tachycardia episodes, which can be
viewed during a patient’s follow-up session. When the maximum number of episodes is exceeded,
the most recent episode replaces the oldest stored episode; however, the rst treated episode will
not be overwritten.
Note: If a spontaneous episode occurs during a patient’s follow-up session
while the SQ-RX device is wirelessly communicating with the
programmer, the episode will not be stored.