QTECH™ PROGRAMMER: OPERATION
The following details are available for each episode:
Treated Episodes
Up to 128 seconds of S-ECG data is stored for each Treated Episode:
Pre-episode S-ECG: Up to 44 seconds
First shock: Up to 24 seconds of pre-shock S-ECG and up to 12 seconds of post-shock
S-ECG
Subsequent shocks: 6 seconds of pre-shock and 6 seconds post-shock S-ECG
Untreated Episodes
An Untreated Episode is dened as any high-rate episode that spontaneously terminates
during the charging process, before a shock is delivered.
Up to 128 seconds of S-ECG data is stored for each Untreated Episode:
Pre-episode S-ECG: 44 seconds of pre-episode S-ECG
Episode S-ECG: Up to 84 seconds of tachycardia S-ECG data
Printing Reports from the Programmer
Printing Reports
Patient reports can be printed before or after a patient session is ended. It is recommended that
a nal report be printed immediately following the implant procedure. There are three patient
reports:
Summary Report
Captured S-ECG Report
Episode Reports
To print patient reports from either an Online or Oine session:
1. Select the Main Menu icon to display the Main Menu screen.
2. Select the Print Reports button to display the Print Reports screen (Figure 37).