User's Manual

Table Of Contents
56
QTECH™ PROGRAMMER: OPERATION
Figure 46: The patient should remain still while the Reference S-ECG is acquired
Capture All Sense Vectors
The Capture All Sense Vectors button on the Utilities screen congures temporary programmer
settings that allow you to capture S-ECGs generated from each of the three sense vectors (Primary,
Secondary, and Alternate). This process takes approximately one minute. The programmer returns
to its original settings conguration after all S-ECGs have been captured.
To capture the three sense vectors:
1. From the Utilities screen (accessible from the Main Menu screen), select the
Capture All Sense Vectors button.
2. The Capturing 12 Second S-ECG screen will appear and display the status of the sense
vector capture process (Figure 47).