QTECH™ PROGRAMMER: OPERATION
2. Select the Reset Beeper button to suspend audible beeping tones triggered by an
alert condition (Figure 48). If the alert condition is not corrected, the audible beeping
tones will be reactivated during the next automatic S-ICD System self-check.
Figure 48: Resetting the Beeper
3. The disable Beeper feature is enabled when ERI is reached or EOL occurs. If desired,
select disable Beeper when ERI or EOL occurs. Beeper tones are permanently
disabled.
Note: If beeper is disabled alert conditions will not trigger audible tones.
Manual Setup
Manual Setup enables the user to perform the electrode integrity test and select the electrode
sensing conguration and gain setting in the SQ-RX device.
1. From the Utilities screen (accessible from the Main Menu screen), select the
Manual Setup button. The Impedance Test screen appears (Figure 49).
2. Select the Test button to perform the electrode integrity test.