User's Manual

Table Of Contents
Implant the leads via the surgical approach chosen.
NOTE: Should lead performance changes occur which cannot be resolved with programming, the lead
may need to be replaced if no adapter is available.
Step D: Take Baseline Measurements
Once the leads are implanted, take baseline measurements. E valuate the lead signals. If performing a pulse
generator replacement procedure, existing leads should be reevaluated, (e.g., signal amplitudes, pacing
thresholds, and impedance). The use of radiography may help ensure lead position and integrity. If testing
results are unsatisfactory, lead system repositioning or replacement may be required.
Connect the pace/sense lead(s) to a pacing system analyzer (PSA).
WARNING: For leads that require the u se of a Connector Tool, use caution handling the lead terminal when
the Connector Tool is not present on the lead. Do not directly contact the lead terminal with any surgical
instruments or electrical connections such as PSA (alligator) clips, ECG connections, forceps, hemostats, and
clamps. This could damage the lead terminal, possibly compromising the sealing integrity and result in loss
of therapy or inappropriate therapy, such as a short within the header.
Pace/sense lead measurements, measured approximately 10 minutes after initial placement (acute) or
during a replacement procedure (chronic), are listed below. Values other than what are s uggested in
the table may be clinically acceptable if appropriate sensing can be documented with the currently
programmed v alues. Consider reprogramming the sensitivity parameter if inappropriate sensing
is observed. Note that the pulse generator measurements may not exactly correlate to the PSA
measurements due to signal ltering.
Table 8. Lead measurements
Pace/ sense lead
(acute)
Pace/ sen se lead
(chronic)
Shocking lead (acute
and chronic)
51