CAUTION: D o not insert a lead into the pulse generator connector w ithout taking the following
precautions to ensure proper lead insertion:
Insert the torque wrench into the preslit depression of the seal plug before inserting the lead into the
port, to release any trapped uid or air.
Visually verify that the setscrew is sufciently retracted to allow insertion. Use the torque wrench
to loosen the setscrew if necessary.
Fully insert each lead into its lead port and then tighten the setscrew onto the terminal pin.
Figure 7. Inserting the torque wrench
d. With the torque wrench in place, fully insert the lead terminal into the lead port. The lead terminal pin
should be clearly visible beyond the connector block when viewed through the side of the pulse generator
header. Place pressure on the lead to maintain its position and ensure that it remains fully inserted
in the lead port.