The High Impedance Limit is nominally set to 2000 , and is programmable between 2000 and 3000 in
250 increments. The Low Impedance Limit is nominally set to 200 , and is programmable between 200
and 500 in 50 increments. Consider the following factors when choosing a value f or the High and Low
Impedance Limits:
For chronic leads, historical impedance measurements for the lead, as well as other electrical
performance indicators such as stability over time
For newly implanted leads, the starting measured impedance value
NOTE: Depending on lead maturation effects, during follow-up testing t he physician may choose to reprogram
the High or Low Impedance Limits.
Pacing dependence of the patient
Recommended impedance range for the lead(s) being used, if available
The Shock Low Impedance Limit is xed at 20 . The Shock High Impedance Limit is nominally set to 125
, and is programmable between 125 and 200 in 25 increments. Consider the following factors when
choosing a value for the High Impedance Limits:
For chronic leads, historical impedance measurements for the lead, as well as other electrical
performance indicators such as stability over time
For newly implanted leads, the starting measured impedance value
NOTE: Depending on lead maturation effects, during follow-up testing the physician may choose to
reprogram the High Impedance Limits.
Recommended impedance range for the lead(s) being used, if available
The impedance value of a high or maximum energy shock impedance test