Magnet Response set to Inhibit Therapy. Once the Patient Triggered Monitor feature has been
triggered by the magnet and an EGM has been stored, or after 60 days have elapsed from the day
that Store EGM was enabled, the Magnet Res ponse programming automatically will be set to Inhibit
Therapy. When this happens, the patient should not apply the magnet because tachyarrhythmia therapy
could be inhibited.
PRECAUTIONS
Clinical Considerations
Pacemaker-mediated tachycardia (PMT). Programming minimum PVARP less than retrograde V–A
conduction may increase the likelihood of a PMT.
Sterilization an d Storage
If package is damaged. The blister t rays and contents are sterilized with ethylene oxide gas before
nal packaging. When the pulse generator and/or lead is received, it is sterile provided the container is
intact. If the packaging is wet, punctured, opened, or otherwise damaged, return the pulse generator
and/or lead to Boston Scientic.
If device is dropped. Do not implant a device which has been dropped while outside of its intact shelf
package. Do not implant a device which has been dropped from a height of m ore than 24 inches (61 cm)
while within its intact shelf package. Sterility, integrity and/or function cannot be guaranteed under these
conditions and the device should be returned to Boston Scientic for inspection.
Storage tem perature and equilibration. Recommended storage temperatures are 0°C–50°C
(32°F–122°F). Allow the device to reach a proper temperature before using telemetry communication
capabilities, programming or implanting the device because temperature ex tremes may affect initial
device function.
•Devicestorage.Store the pulse generator in a clean area away from magnets, kits containing magnets,
and sources of EMI to avoid device damage.
7