User's Manual

Table Of Contents
Do not bend the lead near the lead-header interface. Insert the lead terminal straight into the lead
port. Do not bend the lead near the lead-header interface. Improper insertion can cause insulation
or connector damage.
Absence o f a lead. The absence of a lead or plug in a lead port may affect device performance. If a
lead is not used, be sure to properly insert a plug in the unused port, and then tighten the setscrew
onto the plug.
Electrode connections. Do not insert a lead into the pulse generator connector without taking the
following precautions to ensure proper lead insertion:
Insert the torque wrench into the preslit depression of the seal plug before inserting the lead into the
port, to release any trapped uid or air.
Visually verify that the setscrew is sufciently retracted to allow insertion. Use the torque wrench
to loosen the setscrew if necessary.
Fully insert each lead into its lead port and then tighten the setscrew onto the terminal pin.
•Debrillation lead impedance. If total shocking lead impedance during implant is less than 20 ,verify
the proximal coil is not in contact with the pulse generator surface. A measurement of less than 20 is
an indication of a short somewhere in the system. If repeated measurements show the total shocking
lead impedance is less than 20 , the lead and/or pulse generator may need to be replaced.
Shunting energy. Do not allow any object that is electrically conductive to come into contact with the
lead or device during induction because it may shunt energy, resulting in less energy getting to the
patient, and m ay damage the implanted system.
Do not suture directly over lead. Do not suture directly over the lead body, as this may cause structural
damage. Use the suture sleeve to secure the lead proximal to the venous entry site to prevent lead
movement.
9