User's Manual

Table Of Contents
Fear of shocking while conscious
Fear that shocking capability may be lost
Imagined shocking
Fear of device malfunction
MECHANICAL SPECIFICATIONS
All Extended Longevity (EL) ICD models have a case electrode surface area of 6192 mm². Usable battery
capacity is 1.9 Ah and residual usable battery capacity at Explant is 0.12 Ah for single chamber devices and
0.12 Ah for dual chamber devices. Mechanical specications specic to each model are listed below.
All MINI ICD models have a case electrode surface area of 5487 mm². Usable battery capacity is 1.0 Ah and
residual usable battery capacity at Explant is 0.12 Ah for single chamber devices and 0.12 Ah for dual chamber
devices. Mechanical specications specic to each model are listed below.
Table 2. Mechanical Specications - DYNAGEN Extended Longevity (EL) ICDs
Model Dimensions
WxHxD(cm)
Mass (g) Volume (cm
3
) Connector Type
D150 (VR)
5.37 x 7.36 x 0.99 68.9 29.5 RV: DF4
D151 (VR)
5.37 x 7.79 x 0.99 70.7 31.5
RV: IS-1/DF–1
D152 (DR)
5.37 x 7.68 x 0.99 71.4 31.0
RA:IS-1;RV:DF4
D153 (DR)
5.37 x 7.79 x 0.99 71.0 31.5
RA: IS-1; RV:
IS-1/DF–1
27