The longevity expectations, which account for the energy used during manufacture and storage, apply at the
conditions shown in the table along with the following:
Assumes 60 ppm LRL, ventricular and atrial settings of 2.5 V pacing pulse Amplitude and 0.4 ms pacing
pulse width; RA Imp edance 500 ;sensorsOn.
Projected longevity is calculated assuming 3 maximum energy charging cycles per year, including
automatic capacitor re-forms and therapeutic shocks. For the nal year of device service, an additional
5 charging cycles are assumed to account for additional automatic capacitor re-forms as the device
approaches the Explant indicator. These calculations also assume 3-channel EGM Onset is set to O n,
and that the pulse generator spends 6 months in Storage mode during shipping and storage.
Table 10. Extended Longevity (EL) ICD pulse generator life expectancy estimation (implant to explant)
All Mo dels
ab
Longevity (years) at 500 , 700 ,
and 900 Pacing Impedance (RV)
500 700 900
Pacing VR DR VR DR VR DR
0%
11.7 11.2 11.7 11.2 11.7 11.2
15%
11.5 10.8 11.5 10.9 11.6 10.9
39