User's Manual

Table Of Contents
Table 10. Extended Longevity (EL) ICD pulse g enerator life expectancy estimation (implant to explant)
(continued)
All Mo dels
ab
Longevity (years) at 500 , 700 ,
and 900 Pacing Impedance (RV)
500 700 900
Pacing VR DR VR DR VR DR
50%
11.0 10.0 11.1 10.1 11.2 10.2
100%
10.3 9.0 10.6 9.2 10.8 9.3
a. Assumes ZIP telemetry use for 1 hour at implant time and for 40 minutes annually for in-clinic follow-up checks.
b. Assumes standard use of the LATITUDE Communicator as follows: Daily Device Check on, monthly Full Interrogations
(scheduled remote follow ups, and quarterly patient-initiated interrogations).
Table 11. MINI ICD pulse g enerator life expectancy estimation (implant to explant)
All Mo dels
ab
Longevity (years) at 500 , 700 ,
and 900 Pacing Impedance (RV)
500 700 900
Pacing VR DR VR DR VR DR
0%
5.5
5.3
5.5
5.3
5.5
5.3
40