Table 11. MINI ICD pulse generator life expectancy estimation (implant to explant) (continued)
All Mo dels
ab
Longevity (years) at 500 , 700 ,
and 900 Pacing Impedance (RV)
500 700 900
Pacing VR DR VR DR VR DR
15%
5.4 5.1 5.4 5.1
5.5
5.1
50%
5.24.75.34.75.34.8
100%
4.94.25.04.35.14.4
a. Assumes ZIP telemetry use for 1 hour at implant time and for 40 minutes annually for in-clinic follow-up checks.
b. Assumes standard use of the LATITUDE Communicator as follows: Daily Device Check on, monthly Full Interrogations
(scheduled remote follow ups, and quarterly patient-initiated interrogations).
NOTE: The energy consumption in the longevity table is based upon theoretical electrical principles and
veried via bench testing only.
The pulse generator longevity may increase with a decrease in any of the following:
Pacing rate
Pacing pulse amplitude(s)
Pacing pulse width(s)
Percentage of paced to sensed events
Charging frequency