User's Manual
Table Of Contents
- toc
- Additional Technical Information
- Device Description
- Therapies
- Leads
- PRM System
- Related Information
- Indications and Usage
- Contraindications
- WARNINGS
- General
- Labeling knowledge . Read this manual thoroughly before implanta
- For single patient use only . Do not reuse, reprocess, or rester
- Backup defibrillation protection . Always have external defibril
- Resuscitation availability . Ensure that an external defibrillat
- Separate pulse generator . Do not use this pulse generator with
- Handling
- Programming and Device Operations
- Post-Implant
- Protected environments . Advise patients to seek medical guidanc
- Magnetic Resonance Imaging (MRI) exposure . Do not expose a pati
- Diathermy . Do not subject a patient with an implanted pulse gen
- Ensure PTM is enabled . If desired, ensure that Patient Triggere
- Magnet Response set to Inhibit Therapy . Once the Patient Trigge
- General
- PRECAUTIONS
- Clinical Considerations
- Sterilization and Storage
- If package is damaged . The blister trays and contents are steri
- If device is dropped . Do not implant a device which has been dr
- Storage temperature and equilibration . Recommended storage temp
- Device storage . Store the pulse generator in a clean area away
- Use by date . Implant the pulse generator and/or lead before or
- Implantation
- Expected benefits . Determine whether the expected device benefi
- Evaluate patient for surgery . There may be additional factors r
- Lead compatibility . Prior to implantation, confirm the lead-to-
- Telemetry wand . Make sure a sterile telemetry wand is available
- Line-powered equipment . Exercise extreme caution if testing lea
- Replacement device . Implanting a replacement device in a subcut
- Do not bend the lead near the lead-header interface . Insert the
- Absence of a lead . The absence of a lead or plug in a lead port
- Electrode connections . Do not insert a lead into the pulse gene
- Defibrillation lead impedance . If total shocking lead impedance
- Shunting energy . Do not allow any object that is electrically c
- Do not suture directly over lead . Do not suture directly over t
- Device Programming
- Device communication . Use only the designated PRM and software
- STAT PACE settings . When a pulse generator is programmed to STA
- Pacing and sensing margins . Consider lead maturation in your ch
- Proper programming of the shock vector . If the Shock Vector is
- Programming for supraventricular tachyarrhythmias (SVTs) . Deter
- Adaptive-rate pacing . Rate Adaptive Pacing should be used with
- Ventricular refractory periods (VRPs) in adaptive-rate pacing .
- Shock waveform polarity . For IS-1/DF-1 leads, never change the
- Tachy Mode to Off . To prevent inappropriate shocks, ensure that
- Atrial oversensing . Take care to ensure that artifacts from the
- ATR entry count . Exercise care when programming the Entry Count
- ATR exit count . Exercise care when programming the Exit Count t
- Proper programming without an atrial lead . If an atrial lead is
- Sensing adjustment . Following any sensing range adjustment or a
- Patients hear tones coming from their devices . Patients should
- Patient use of patient triggered monitor . Determine if the pati
- Patient initiate stored EGM . Consider having the patient initia
- Environmental and Medical Therapy Hazards
- Hospital and Medical Environments
- Mechanical ventilators . During mechanical ventilation, respirat
- Conducted electrical current . Any medical equipment, treatment,
- Internal defibrillation . Do not use internal defibrillation pad
- External defibrillation . It can take up to 15 seconds for sensi
- Lithotripsy . Extracorporeal shock wave lithotripsy (ESWL) may c
- Ultrasound energy . Therapeutic ultrasound (e.g., lithotripsy) e
- Electrical interference . Electrical interference or “noise” fro
- Radio frequency (RF) interference . RF signals from devices that
- Central line guidewire insertion . Use caution when inserting gu
- Home and Occupational Environments
- Follow-up Testing
- Explant and Disposal
- Supplemental Precautionary Information
- Potential Adverse Events
- Mechanical Specifications
- Table 2. Mechanical Specifications - DYNAGEN Extended Longevity
- Table 3. Mechanical Specifications - DYNAGEN MINI ICDs
- Table 4. Mechanical Specifications - INOGEN Extended Longevity (
- Table 5. Mechanical Specifications - INOGEN MINI ICDs
- Table 6. Mechanical Specifications - ORIGEN Extended Longevity (
- Table 7. Mechanical Specifications - ORIGEN MINI ICDs
- Items Included in Package
- Symbols on Packaging
- Characteristics as Shipped
- X-Ray Identifier
- Federal Communications Commission (FCC)
- Pulse Generator Longevity
- Warranty Information
- Product Reliability
- Patient Counseling Information
- Lead Connections
- CAUTION: Prior to implantation, confirm the lead-to-pulse genera
- Figure 2. Lead connections and setscrew locations, RA: IS-1, RV:
- Figure 3. Lead connections and setscrew locations, RV: DF4-LLHH
- Figure 4. Lead connections and setscrew locations, RA: IS-1, RV:
- Figure 5. Lead connections and setscrew locations, RV: IS-1/DF-1
- Implanting the Pulse Generator
- CAUTION: The absence of a lead or plug in a lead port may affect
- CAUTION: Do not suture directly over the lead body, as this may
- WARNING: For leads that require the use of a Connector Tool, use
- Table 12. Lead measurements
- WARNING: For leads that require the use of a Connector Tool, use
- WARNING: Do not contact any other portion of the DF4–LLHH or DF4
- CAUTION: For IS-1/DF-1 leads, never change the shock waveform po
- CAUTION: Do not insert a lead into the pulse generator connector
- Figure 6. Inserting the torque wrench
- CAUTION: Insert the lead terminal straight into the lead port. D
- CAUTION: The absence of a lead or plug in a lead port may affect
- CAUTION: To prevent inappropriate shocks, ensure that the pulse
- CAUTION: Take care to ensure that artifacts from the ventricles
- CAUTION: If total shocking lead impedance during implant is less
- CAUTION: To prevent inappropriate shocks, ensure that the pulse
- WARNING: Always have external defibrillation equipment available
- Induce the Patient’s Arrhythmia
- Perform the Induction
- Determine DFT
- Bidirectional Torque Wrench
- Loosening Stuck Setscrews
- Follow Up Testing
- For additional technical reference guides, go to
- Additional Technical Information

Selection of lead configuration and specific surgical procedures is a matter of professional judgment. The
following leads are available for use with the pulse generator depending on the device model.
• ENDOTAK endocardial cardioversion/defibrillation and pacing lead system
• Ventricular endocardial bipolar lead
• Atrial bipolar lead
• Superior vena cava lead coupled with a ventricular patch lead
• Two-patch epicardial leads configuration
CAUTION: The absence of a lead or plug in a lead port may affect device performance. If a lead is not used,
be sure to properly insert a plug in the unused port, and then tighten the setscrew onto the plug.
CAUTION: Do not suture directly over the lead body, as this may cause structural damage. Use the suture
sleeve to secure the lead proximal to the venous entry site to prevent lead movement.
Whichever lead configuration is used for both pacing/sensing and defibrillating, several considerations and
cautions should be heeded. Factors such as cardiomegaly or drug therapy may necessitate repositioning of the
defibrillating leads or s ubstituting one lead for another to facilitate arrhythmia conversion. In some instances,
no lead configuration may be found that provides reliable arrhythmia termination at energy levels available from
the pulse generator. Implantation of the pulse generator is not recommended in these cases.
Implant the leads via the surgical approach chosen.
NOTE: Should lead performance changes occur which cannot be resolved with programming, the lead
may need to be replaced if no adapter is available.
Step D: Take Baseline Measurements
Once the leads are implanted, take baseline measurements. Evaluate the lead signals. If performing a pulse
generator replacement procedure, existing leads should be reevaluated, (e.g., signal amplitudes, pacing
50