User's Manual
Table Of Contents
- toc
- Additional Technical Information
- Device Description
- Therapies
- Leads
- PRM System
- Related Information
- Indications and Usage
- Contraindications
- WARNINGS
- General
- Labeling knowledge . Read this manual thoroughly before implanta
- For single patient use only . Do not reuse, reprocess, or rester
- Backup defibrillation protection . Always have external defibril
- Resuscitation availability . Ensure that an external defibrillat
- Separate pulse generator . Do not use this pulse generator with
- Handling
- Programming and Device Operations
- Post-Implant
- Protected environments . Advise patients to seek medical guidanc
- Magnetic Resonance Imaging (MRI) exposure . Do not expose a pati
- Diathermy . Do not subject a patient with an implanted pulse gen
- Ensure PTM is enabled . If desired, ensure that Patient Triggere
- Magnet Response set to Inhibit Therapy . Once the Patient Trigge
- General
- PRECAUTIONS
- Clinical Considerations
- Sterilization and Storage
- If package is damaged . The blister trays and contents are steri
- If device is dropped . Do not implant a device which has been dr
- Storage temperature and equilibration . Recommended storage temp
- Device storage . Store the pulse generator in a clean area away
- Use by date . Implant the pulse generator and/or lead before or
- Implantation
- Expected benefits . Determine whether the expected device benefi
- Evaluate patient for surgery . There may be additional factors r
- Lead compatibility . Prior to implantation, confirm the lead-to-
- Telemetry wand . Make sure a sterile telemetry wand is available
- Line-powered equipment . Exercise extreme caution if testing lea
- Replacement device . Implanting a replacement device in a subcut
- Do not bend the lead near the lead-header interface . Insert the
- Absence of a lead . The absence of a lead or plug in a lead port
- Electrode connections . Do not insert a lead into the pulse gene
- Defibrillation lead impedance . If total shocking lead impedance
- Shunting energy . Do not allow any object that is electrically c
- Do not suture directly over lead . Do not suture directly over t
- Device Programming
- Device communication . Use only the designated PRM and software
- STAT PACE settings . When a pulse generator is programmed to STA
- Pacing and sensing margins . Consider lead maturation in your ch
- Proper programming of the shock vector . If the Shock Vector is
- Programming for supraventricular tachyarrhythmias (SVTs) . Deter
- Adaptive-rate pacing . Rate Adaptive Pacing should be used with
- Ventricular refractory periods (VRPs) in adaptive-rate pacing .
- Shock waveform polarity . For IS-1/DF-1 leads, never change the
- Tachy Mode to Off . To prevent inappropriate shocks, ensure that
- Atrial oversensing . Take care to ensure that artifacts from the
- ATR entry count . Exercise care when programming the Entry Count
- ATR exit count . Exercise care when programming the Exit Count t
- Proper programming without an atrial lead . If an atrial lead is
- Sensing adjustment . Following any sensing range adjustment or a
- Patients hear tones coming from their devices . Patients should
- Patient use of patient triggered monitor . Determine if the pati
- Patient initiate stored EGM . Consider having the patient initia
- Environmental and Medical Therapy Hazards
- Hospital and Medical Environments
- Mechanical ventilators . During mechanical ventilation, respirat
- Conducted electrical current . Any medical equipment, treatment,
- Internal defibrillation . Do not use internal defibrillation pad
- External defibrillation . It can take up to 15 seconds for sensi
- Lithotripsy . Extracorporeal shock wave lithotripsy (ESWL) may c
- Ultrasound energy . Therapeutic ultrasound (e.g., lithotripsy) e
- Electrical interference . Electrical interference or “noise” fro
- Radio frequency (RF) interference . RF signals from devices that
- Central line guidewire insertion . Use caution when inserting gu
- Home and Occupational Environments
- Follow-up Testing
- Explant and Disposal
- Supplemental Precautionary Information
- Potential Adverse Events
- Mechanical Specifications
- Table 2. Mechanical Specifications - DYNAGEN Extended Longevity
- Table 3. Mechanical Specifications - DYNAGEN MINI ICDs
- Table 4. Mechanical Specifications - INOGEN Extended Longevity (
- Table 5. Mechanical Specifications - INOGEN MINI ICDs
- Table 6. Mechanical Specifications - ORIGEN Extended Longevity (
- Table 7. Mechanical Specifications - ORIGEN MINI ICDs
- Items Included in Package
- Symbols on Packaging
- Characteristics as Shipped
- X-Ray Identifier
- Federal Communications Commission (FCC)
- Pulse Generator Longevity
- Warranty Information
- Product Reliability
- Patient Counseling Information
- Lead Connections
- CAUTION: Prior to implantation, confirm the lead-to-pulse genera
- Figure 2. Lead connections and setscrew locations, RA: IS-1, RV:
- Figure 3. Lead connections and setscrew locations, RV: DF4-LLHH
- Figure 4. Lead connections and setscrew locations, RA: IS-1, RV:
- Figure 5. Lead connections and setscrew locations, RV: IS-1/DF-1
- Implanting the Pulse Generator
- CAUTION: The absence of a lead or plug in a lead port may affect
- CAUTION: Do not suture directly over the lead body, as this may
- WARNING: For leads that require the use of a Connector Tool, use
- Table 12. Lead measurements
- WARNING: For leads that require the use of a Connector Tool, use
- WARNING: Do not contact any other portion of the DF4–LLHH or DF4
- CAUTION: For IS-1/DF-1 leads, never change the shock waveform po
- CAUTION: Do not insert a lead into the pulse generator connector
- Figure 6. Inserting the torque wrench
- CAUTION: Insert the lead terminal straight into the lead port. D
- CAUTION: The absence of a lead or plug in a lead port may affect
- CAUTION: To prevent inappropriate shocks, ensure that the pulse
- CAUTION: Take care to ensure that artifacts from the ventricles
- CAUTION: If total shocking lead impedance during implant is less
- CAUTION: To prevent inappropriate shocks, ensure that the pulse
- WARNING: Always have external defibrillation equipment available
- Induce the Patient’s Arrhythmia
- Perform the Induction
- Determine DFT
- Bidirectional Torque Wrench
- Loosening Stuck Setscrews
- Follow Up Testing
- For additional technical reference guides, go to
- Additional Technical Information

2. Verify magnet function and telemetry to ensure the pulse generator is within acceptable range.
3. Program the appropriate parameters and change the pulse generator Tachy Mode to Monitor + Therapy.
4. Perform the induction using the programmer.
Determine DFT
Defibrillation threshold (DFT) testing should be performed at implant to ensure adequate safety margins for the
shock energy (safety margin = device maximum shocking energy minus DFT). The DFT is the minimum energy
tested that reliably converts VF or polymorphic VT in a series of arrhythmia trials with decreasing energy steps.
To determine DFT, induce VF (or PVT or ventricular flutter if VF is not inducible). Attempt to convert the
arrhythmia with a 31-J shock; if conversion is successful, then induce again and attempt to convert the
arrhythmia at 29 J. Continue in this manner, decreasing the energy level until failure to convert VF occurs or 3 J
is successful. If the initial conversion at 31 J is unsuccessful, reposition the lead, or reverse polarity, or add an
additional lead. The energy level that last successfully converted the arrhythmia is the DFT. It is a matter of
clinical judgment as to what constitutes a demonstration of reliable conversion.
Shocks intended for VF therapy should be programmed with a 10 J safety margin above the shock energy
level that the physician determines is required for successful VF conversion. In some situations an alternative
safety margin may be determined by the physician to be adequate above the shock energy level required
for successful VF conversion.
NOTE: Since the results of any single test may be an example of statistical variation, a one-time conversion of
a rhythm disturbance at a particular energy level does not guarantee or assure that the energy level is reliable
for conversion. As a safety margin, it is recommended that you perform the conversion test at the DFT two
times if the DFT is 31 J, or one time if the DFT is 29 J with no failures to convert.
Step J: Implant the Pulse Generator
1. Program the Tachy Mode to Off.
62