WARNINGS
General
Labeling knowledge. Read this manual thoroughly before implantation to avoid damage to the pulse
generator and/or lead. Such damage can result in patient injury or death.
For single patient use only. Do not reuse, reprocess, or resterilize. Reuse, reprocessing, or
resterilization may compromise the structural integrity of the device and/or lead to device failure which, in
turn, may result in patient injury, illness, or death. Reuse, reprocessing, or resterilization may also create
a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not
limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the
device may lead to injury, illness, or death of the patient.
Backup debrillation p rotection. Always have external debrillation equipment available during
implant and electrophysiologic t esting. If not terminated in a timely fashion, an induced ventricular
tachyarrhythmia can result in the patient’s death.
Resuscitation availability. Ensure that an external debrillator and medical personnel skilled in CPR are
present during post-implant device testing should the patient require external rescue.
Separate pulse generator. Do not use this pulse generator with another pulse generator. This
combination could cause pulse generator interaction, resulting in patient injury or a lack of therapy delivery.
Handling
Avoid shock during han dling. Program the pulse generator Tachy Mode(s) to Off during implant,
explant, or postmortem procedures to avoid inadvertent high voltage shocks.
Do not kink leads. Do not kink, twist, or braid the lead with other leads as doing so could cause lead
insulation abrasion damage or conductor damage.
5