Sensing adjustment. Following any sensing range adjustment or any modication of the sensing lead,
always verify appropriate sensing. Programming Sensitivity to the highest value (lowest sensitivity) may
result in delayed detection or undersensing of cardiac activity. Likewise, programming to the lowest value
(highest sensitivity) may result in oversensing of non-cardiac signals.
Patients hear tones coming from their devices. Patients should be advised to contact their physician
immediately if they hear tones coming from their device.
Patient use of patient triggered monitor. Determine if the patient is capable of activating this feature
prior to being given the magnet and prior to enabling Patient Triggered Monitor. Remind the patient to
avoid strong magnetic elds so the feature is not inadvertently triggered.
Patient initiate stored EGM. Consider having the patient initiate a stored EGM at the time Patient
Triggered Monitor is enabled to assist with patient education and feature validation. Verify the activation
of the feature on the Arrhythmia Logbook screen.