User's Manual

Table Of Contents
Hospital and Medical Environments
Mechanical ventilators. During mechanical ventilation, respiration-based trending may be misleading;
therefore, the Respiratory Sensor should be programmed to Off.
Conducted electrical current. Any medical equipment, treatment, therapy, or diagnostic test that
introduces electrical current into the patient has the potential to interfere with pulse generator function.
External patient monitors (e.g., respiratory m onitors, surface ECG monitors, hemodynamic
monitors) may interfere with the pulse generator’s impedance-based diagnostics (e.g., shock
lead impedance measurements, Respiratory Rate trend). To resolve suspected interactions with
Respiratory Sensor-based diagnostics, deactivate the pulse generator’s Respiratory Sensor by
programming it to Off.
Medical therapies, treatments, and diagnostic tests that use conducted electrical current (e.g.,
TENS, electrocautery, electrolysis/thermolysis, electrodiagnostic testing, electromyography, or
nerve conduction studies) may interfere with or damage the pulse generator. Program the device
to Electrocautery Protection Mode prior to the treatment, and monitor device performance during
the treatment. After the treatment, verify pulse generator function ("Post-Therapy P ulse Generator
Follow Up" on page 19).
•Internaldebrillation. Do not use internal debrillation paddles or catheters unless the pulse generator
is disconnected from the leads because the leads may shunt energy. This could result in injury to the
patient and damage to the implanted system.
15