User's Manual

Table Of Contents
Reviewing c linical events and fault codes
Reviewing the Arrhythmia Logbook, including stored electrograms (EGMs)
Reviewing real-time EGMs
Testing the leads (threshold, amplitude, and impedance)
Performing a manual capacitor re-formation
Reviewing respiratory sensor-based diagnostics
Verifying battery status
Programming any permanent brady parameter to a new value and then reprogramming it back to t he
desired value
Programming the Tachy Mode to a new value and then reprogramming it back to the desired value
Saving all patient data
Verifying the appropriate nal programming prior to allowing the patient to leave the clinic
Transcutaneous Electrical Nerve Stimulation (TENS)
CAUTION: TENS involves passing electrical current through the body, and may interfere with pulse generator
function. If TENS is medically necessary, evaluate the TENS therapy settings for compatibility with the pulse
generator. The following guidelines may reduce the likelihood of interaction:
Place the TENS electrodes as close together and as far away from the pulse generator and leads
as possible.
Use the lowest clinically-appropriate TENS energy output.
Consider cardiac monitoring during TENS use, especially for pacemaker-dependent patients.
20