Table 7. Pulse generator life expectancy estimation (implan t to explant)
All Models
ab
Pacing Amplitude
Longevity (years) at 500 and 700
Pacing Impedance (RV and LV)
RA/RV
LV
500 700
2.5 V 3.0 V 8.1 8.6
2.5 V 3.5 V 7.6 8.2
3.5 V 3.5 V 6.8
7.5
3.5 V 5.0 V
5.7
6.5
a. Assumes ZIP wandless telemetry use for 3 hours at implant time and for 40 minutes annually for in-clinic follow-up checks.
b. Assumes standard u se of the LATITUDE Communicator as follows: Daily Device Check on, monthly Full Interrogations
(scheduled remote follow ups, and quarterly patient-initiated interrogations).
NOTE: The energy consumption in the longevity table is based upon theoretical electrical principles and
veried via bench testing only.
The pulse generator longevity may increase with a decrease in any of the following:
Pacing rate
Pacing pulse amplitude(s)
Pacing pulse w idth(s)
Percentage of paced to sensed events
Charging frequency