User's Manual

Table Of Contents
Table 8. Lead measurements (continued)
Pace/ sense lead
(acute)
Pace/ sense lead
(chronic)
Shocking lead (acute
and chronic)
R-Wave Amplitude
ab
> 5 mV > 5 mV > 1.0 mV
P-Wave Amplitude
ab
> 1.5 mV > 1.5 mV
R-Wave Duration
bcd
< 100 ms < 100 ms
Pacing Threshold (right
ventricle)
< 1.5 V endocardial
< 2.0 V epicardial
< 3.0 V endocardial
<3.5Vepicardial
Pacing Threshold (left
ventricle)
< 2.5 V coronary venous
< 2.0 V epicardial
< 3.5 V coronary venous
<3.5Vepicardial
Pacing Threshold
(atrium)
< 1.5 V endocardial < 3.0 V endocardial
Lead impedance (at 5.0
V and 0.5 ms atrium and
right ventricle)
e
> programmed Low
Impedance Limit
(200–500 )
< programmed High
Impedance Limit
(2000–3000 )
> programmed Low
Impedance Limit
(200–500 )
< programmed High
Impedance Limit
(2000–3000 )
>20
<programmedHigh
Impedance Limit
(125–200 )
52